![EDAN F15 Series Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/edan/f15-series/f15-series_service-manual_2357161034.webp)
F15 Series Fetal & Maternal Monitor Service Manual
Principle Introduction
- 23 -
Serial Port
8 serial port ways
Touch Screen Interface
1 interface.
USB
1 standard USB HOST interface, 1 standard DEVICE interface (1.1
specification)
HDMI
1 HDMI output port
LCD
eDP
interface,
1920*1080@60Hz
(highest resolution)
Loudspeaker Interface
1 interface.
Stepping Motor Interface 1 interface.
Printer Head Interface
1 interface.
Out of Paper Checking
Interface
1 interface.
Paper Darwer Checking
Interface
1 interface.
Power Consumption
8.5W
Size
180 mm×155 mm
Operating System
Linux
6.1.2 Bottom Interface Board
The bottom interface board locates at the left rear side of the lower case. It fulfills the
following function: communicate with external USB interface, network interface, HDMI, serial
port (RS-485*1
,
FTS-3 serial port*2
)
and bridging signals of transducer accessory serial port,
MARK, EXT to the main control board.
Содержание F15 Series
Страница 1: ......
Страница 15: ...F15 Series Fetal Maternal Monitor Service Manual Safety Guidance 4 42 DO NOT STEP ON ...
Страница 51: ...F15 Series Fetal Maternal Monitor Service Manual Modules Malfunction Verification 40 Main control board TP61 ...
Страница 75: ...F15 Series Fetal Maternal Monitor Service Manual Periodic Tests 64 accepted ...
Страница 80: ......