![EDAN F15 Series Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/edan/f15-series/f15-series_service-manual_2357161069.webp)
F15 Series Fetal & Maternal Monitor Service Manual
Disassembling the Monitor
- 58 -
CAUTION
The thermal printhead is a fragile part. Handle it with care.
9.4.11 Replacing the AP Module (wireless model only)
To remove the AP module,
1)
Separate the upper assembly and the lower assembly.
2)
Remove all the cables on the AP module.
3)
Remove the screw on the AP module.
To replace the AP module,
Replace the AP module in a reverse procedure. Refer to the following illustration when connecting
wires. Ensure they are all correctly connected.
9.4.12 Replacing the Wireless Transducer Bracket (wireless model only)
To remove the wireless transducer holder bracket,
1)
Remove the two screws securing the wireless transducer holder bracket.