F15 Series Fetal & Maternal Monitor Service Manual
Principle Introduction
- 27 -
Measurement Time Span
120 Seconds (Maximum)
Software Overpressure Protection
(297 ± 3) mmHg
Hardware Overpressure Protection
(320 ± 10)mmHg [(41.3±1.3) kPa]
(
2
)
SpO2
(
EDAN
)
SpO
2
Measurement Range
50% ~ 100%
Resolution
1%
Accuracy
70% ~ 100%
:
±
2%
50% ~ 69%
:
Not Defined
Pulse
Measurement Range
30 bpm ~ 240 bpm
Resolution
1 bpm
Accuracy
±
3 bpm
3
)
SpO2
(
Nellcor
)
SpO
2
Measurement Range
50% ~ 100%
Resolution
1%
Accuracy
70% ~ 100%
:
±
2%
50% ~ 69%
:
Not Defined
Pulse
Measurement Range
30 bpm ~ 240 bpm
Resolution
1 bpm
Accuracy
±
3 bpm
4
)
TEMP
Channel Qty.
1
Measurement Range
0 ºC ~ +50.0 ºC (+32ºF ~ +122ºF)
Resolution
0.1 ºC (±0.18ºF)
Accuracy
(Sensor inaccuracy not included)
+25ºC ~ +45ºC (+77ºF ~ +113ºF): ± 0.2ºC (±0.36ºF)
Other measurement range: ± 0.3ºC