Page 49
Model TC80
F
AULT
S
TATUS
Bit #
bit meaning
0
Memory Error, either a Program Flash, RAM or NVM RAM checksum error has occurred
1
Input Voltage Out Of Tolerance
2
The On Board +12V is Out of Tolerance
3
The On Board +3.3V is Out of Tolerance
4
The Transmitter has lost communication link with the Sensor
5
There is no Sensor connected
6
Sensor Calibration Failed
7
Relay 1 on-time expired
8
Relay 2 on-time expired
9
Relay 3 on-time expired
10
Sentinel Error (useable life has expired)
11
Sentinel Poisoned
12
Membrane Error
13
NU
14
NU
15
NU
W
ARNING
S
TATUS
Bit #
bit meaning
0
The Sensor has changed from previously connect Sensor
1
Not Used (NU)
2
NU
3
NU
4
NU
5
NU
6
NU
7
NU
8
NU
9
NU
10
NU
11
NU
12
NU
13
NU
14
NU
15
NU
S
ENSOR
T
YPE
Data
Meaning
Decimal Hexadecimal
Chemical
Sensor Type
Measurement
Units
0
0000
Unknown
Chemical
None
None
1
0001
Ammonia
mV
ppm
2
0002
Ammonium
mV
ppm
3
0003
Bromide
mV
ppm
4
0004
Calcium
mV
ppm
5
0005
Chloride
mV
ppm
6
0006
Conductivity
Conductivity
S
7
0007
Cupric
mV
ppm
8
0008
Cyanide
mV
ppm
9
0009
DO
mV
ppm
10
000A
DO
mV
% saturation
11
000B
DO
mV
mg/L