Figure 3.2
3.03
I
NFLUENCES ON THE
M
EASUREMENT
T
EMPERATURE
The Total Chlorine Sensor (TCS) digitally outputs a temperature corrected value to the transmitter. The
temperature sensor is located in the TCS sensor and it has a response time of several minutes. Rapid changes of
temperature will introduce an error until the sensor has equilibrated to the new temperature. Calibration should
be done close to the process temperature for the highest accuracy.
0
20
40
60
80
100
0
5
10
15
20
25
30
% output
Flowrate, gal/hr
TCS Flow dependence