Page 33
Model TC80
4.1
P
H
C
ALIBRATION
P
ROCEDURES
AUTO Calibration recognizes pH 4.01, pH 7.00 and pH 10.00 buffer solutions for automatic, temperature
compensated calibrations. Any calibration solutions can be used but the pH value will have to be entered
manually. Follow the steps below to accomplish a pH calibration. Example uses pH 7.00 and pH 4.01 buffers.
4.1.1
AUTO
C
AL
USING P
H
4.01,
7.00,
10.00
BUFFERS
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press
HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Press
CAL
Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)
Press Yes/
NO
Place Sensor in CAL Solution ( use pH 7.00 buffer)
Press
AUTO
then
CAL 1
STABILIZING, 7.00 pH xx.x mV, 7.00 pH corrected Accept Cal 1?
Press
YES
CAL1 Value 7.00 pH, Continue to CAL2? Move sensor to 4.01 pH buffer solution
Press
YES
STABILIZING, 4.00 pH xxx.x mV, 4.00 pH corrected Accept Cal?
Press
YES
OFFSET: 7.00 pH xx.x mV, SLOPE: 59.16 mV/pH (data written to Log)
Press
OK
Calibration complete
Press
HOME
Hold is ON
Press
HOLD
Turn off Hold
Press
EXIT
Main Display
4.1.2
AUTO
C
AL
USING OTHER P
H
BUFFERS
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press
HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration
Press
CAL
Is this a new Sensor? (Yes erases CAL Log in INFO, NO adds CAL to existing Log)
Press
YES
/NO
Place Sensor in CAL Solution
Press
AUTO then CAL 1
STABILIZING, 6.86 pH 8.2 mV, 7.00 pH corrected Accept Cal?
Press
NO
Enter CAL 1 Value
Press
▲
▼
NEXT
6.86 pH (use arrows and NEXT to enter pH Buffer value)
Press
OK
6.86 pH, 8.2 mV, Accept this Value
Press
YES
CAL 1 Value 6.86 pH, Continue to CAL 2? (Place Sensor in 2
nd
calibration buffer)
Press
YES
STABILIZING, 9.18pH 135.6 mV, 10.00 pH corrected Accept Cal?
Press
NO
Enter CAL 2 Value
Press
▲
▼
NEXT
9.18 pH (use arrows and NEXT to enter pH Buffer value)
Press
OK
9.18 pH, 135.6 mV, Accept this Value
Press
YES
OFFSET: 6.86 pH 8.2 mV, SLOPE: 59.16 mV/pH (data written to Log)
Press
OK
Calibration complete
Press
HOME
Hold is ON
Press
HOLD
Turn off Hold
Press
EXIT
Main Display
4.1.3
S
TANDARDIZE
Leave the sensor in the process solution, take a grab sample from the process and determine the pH or place
sensor in a calibration standard solution.
Action
Prompt
Double Press any Button
MENU HOME, Hold is OFF
Press
HOLD
Hold freezes 4-20 mA Output and locks Alarm Relays during Calibration