Page 13
Model TC80
1.1
FEATURES
Panel Mounted System
, Easy Installation
Plumb and Play Design
, Ready to Use
Automatic pH Compensation
, No Expensive Reagents to
mix or spill
Automatic Flow Control
, Eliminates Pressure Regulators
and Rotameters
T80 Transmitter Capability
, Dual Measurements,24VDC or
110/220 VAC Power, Graphical Plots
Compliant with EPA Method 334.0
1.2
SPECIFICATIONS
1.2.1
S
ENSORS AND
F
LOW
T
RAIN
Chlorine Sensor:
Digital protocol, Potentiostatic, Gold cathode/Silver-Silver Halide anode, 316L SS counter electrode
pH Sensor:
Digital S80 protocol, 316L stainless steel body with replaceable electrode cartridge
Measurement Range:
Chlorine: 0.05 to 20 ppm (High Range)
0.005 to 2.00 ppm (Low Range)
pH: 4 to 14 pH
Operating Temperature:
0° C to 50° C (32° F to 122° F)
Min/Max Flow:
38 L/hr. to 300 L/hr. (10 gal/hr. to 80 gal/hr.)
Wetted Materials:
PVC, PP, PVDF, PTFE, Glass, 316 SS
Process Connections:
Input ¼” FNPT with barb fitting, Drain ¾” FNPT
Response Time:
T90 in 2 minutes
Electrolyte Life:
3 to 6 months
1.2.2
TC80
A
NALYZER
Measurements:
Chlorine: 0.00 ppb to 22.00 ppm
pH: 0.00 to 14.00 pH
pH Compensation of Total Chlorine:
pH 4 - 12
Display:
128 x 64 pixels (2.75” x 1.5”) LCD, Black on Grey background, Blue on White background with LED backlight on
Outputs:
(1) 4-20 mA for Total Chlorine set to Sensors Range