Page 25
Model TC80
o
Disable
, Inactivates relay and removes the
relay button from the HOME Screen display.
HOLD,
Freezes outputs at
current value and locks relays in
their current state.
Hold Timeout,
Removes HOLD after a certain period
of time, default setting: No Timeout, selections include 15 minutes, ½ hour, 1
hour
o
SERIAL
MODBUS configure serial output,
ADDRESS,
enter address: 001 to 247
BAUD,
Choose baud rate, default 9600
FORMAT,
set serial data format, default value: 8N1, 8 bit, no
parity bit, 1 stop bit
o
PASSWD
Enter 4 character password to protect access to MENU Level, CAL Menu, CONFIG
Menu and SIM Menu (simulate). Each level can be turned ON or OFF and can have a unique
password.
MENU
ON/OFF _ _ _ _ Locks Main Menu
CAL
ON/OFF _ _ _ _ Locks CAL and CONFIG
CONFIG
ON/OFF _ _ _ _ Locks CONFIG
SIM
ON/OFF _ _ _ _ Locks SIM and CONFIG
SENSOR
enters the sensor configuration menu.
o
Choose SENSOR 1 or 2
TYPE,
Allows T80 transmitter to configure the S80 sensor. For
use only when switching the measurement electrode type in
an S80 sensor, i.e. for a pH electrode to a pION electrode.
Select Sensor Type: pH, ORP, DO
2
, NH
3
, NH
4
+
, Br
-
, Ca
++
, Cl
-
,
Conductivity, Resistivity, Cu
++
, CN
-
, F
-
, NO
3
-
, K
+
, Ag
+
, Na
+
, S
--
T COMP,
Enter % temperature compensation per degree: pH,
0.33%, Total Chlorine 0%,
o
COMP
Dual Channel Only, Sets compensation type:
Dissociation
(pKa),
NH
4
+
, Free Chlorine, HF, S
-2
,
Interference
, X ppm Sensor 2 = 1 ppm
Sensor 1,
Percentage
% change per pH, Total Chlorine -5%.
o
Qty of SENSORS,
Choose 1 or 2
Load Default
resets all Menus to factory default configuration.
Dampen
,
sets the number of measurements averaged for the displayed PV
3.2.4
INFO
(I
NFORMATION
M
ENU
)
The Information Menu provides two choices,
Transmitter Screen
, details the Name, Power type, Serial #, Firmware version
and the output configuration(s).
Sensor Screen
, details the Name, Part #, Serial # and three sets of Calibration data.
COMP
, displays the pKa, the sensor affected and the dissociation Factor,
3.2.5
SIM
(S
IMULATION
M
ENU
)
The Simulation menu allows the Input or Output signals to be simulated.