Page 27
Model TC80
3.3.1
C
ONFIGURE
4-20
M
A
OUTPUT RANGE
Double press any key except the HOME key to enter the HOME Menu. Follow
the path below to set the 4-20 mA range.
HOME Menu → Press CONFIG → XMTR → OUTPUT → 4-20 (1)(2) → PV or
TEMP
Press CHANGE to enter New Values.
Choose 4 mA value, press OK
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK→Back
Choose 20 mA value, press OK,
Enter value using
▲
or
▼
and NEXT to move to the next digit, press OK→Back
Press BACK to return to the CONFIGURE 4-20 mA screen or HOME to return to
the HOME Menu screen.
3.3.2
C
ONFIGURE
4-20
M
A
F
AULT
C
ONDITION AND
C
AL
In the CONFIGURE 4-20 mA screen, Press
MORE
→
FAULT
or
Choose
Low Fault 3.5 mA
or
Hi Fault 22 mA
or
NONE
, (default setting
NONE
),
Press OK
Press BACK →
CAL
, (enter PW 0000) connect DVM to 4-20 mA line, Press 4.00
mA then adjust value to the DVM reading, Press 20.00 mA and adjust value to
the DVM reading. The 4-20 mA output is calibrated.
3.3.3
C
ONFIGURE
A
LARM
R
ELAYS
(R
ELAYS
O
PTIONAL
)
HOME Menu → Press CONFIG → XMTR → OUTPUT → RELAYS→RLY1
Choose the
ALARM
,
TIMER
,
FAULT
or
DISABLE
mode for Relay 1
ALARM
Displays:
o
SET POINT ON
: The Process Variable Value that activates the relay.
o
EXPIRATION
: Enter a time that should not be exceeded before the PV
should have changed enough to activate the OFF set point. At the
Expiration time the relay is deactivated and a Fault condition is
initiated. Fault: Relay 1 Time expired: Cause: Loss of reagent, failed
sensor ….
o
Delay ON
: The amount of time the PV must remain above/below the
set point before the relay activates.
o
SET POINT OFF
: The Value of the process variable that deactivates the
relay.
SET POINT OFF > Set Point → Low Set Point
SET POINT OFF < Set Point → Hi Set Point
o
Delay OFF
: The amount of time the PV must remain above/below the
hysteresis point before the relay deactivates.
o
STATE
:
Energize
(relay is energized on activation)/
De-energize
(relay
is de-energized on activation)
TIMER
activates the relay periodically for a specific duration, user configured
period and duration
FAULT
sets the relay condition to a de-energize state and NC relay closes in
response to a Fault condition or power failure.