MGR10 A/B/C - Revision : I (for 7.0 software revision)
70
Settling Algorithm Group
SENSe:SETTling:STATe <ON|OFF> or <1|0>
Enables or disables the Settling facility.
SENSe:SETTling:STATe?
Returns “0” (OFF) or “1” (ON).
SENSe:SETTling:COUNt <No. of readings>
Sets the maximum number of readings to be taken (2 - 999).
SENSe:SETTling:COUNt?
Returns the maximum number of readings selected.
SENSe:SETTling:LIMit <No. of digits>
Sets the maximum number of display digits by which two consecutive readings may differ, for the
measurement to be defined as ‘settled’.
SENSe:SETTling:LIMit?
Returns the maximum number of display digits selected.
Selecting the Settling facility will:
Deselect the Filter facility.
Change Average (AVE) Current mode to Positive (+I) Current mode.
Cause an error message if in Fast mode.
Selecting Filter on, Average Current or Fast mode will automatically deselect the Settling facility.
Trigger & Measure Group
These commands trigger measurements, and allow the results to be output to the bus.
INITiate
This initiates single measurements. The Measurement Available bit (8) of the Operation Condition
Register is set on completion. The reading is stored internally and is not sent to the output buffer.
Use FETCh? to transfer the value to the output buffer and clear the Measurement Available bit.
This command is not available when the CONTinuous triggering mode is on, nor when
DATAlogging mode is on, in which cases, an “execution error” will occur.
*TRG
This is identical to the INITiate command
FETCh[:FRESistance|TEMPerature|TCOMpensate]?
Returns the value of the last measurement made. This is normally used in conjunction with the
INITiate command. If CONTinuous triggering is “off”, then INITiate followed by FETCh? will be
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...