MGR10 A/B/C - Revision : I (for 7.0 software revision)
53
6.2 RS232 and IEEE488 PROGRAMMING INTERFACES :
Two different types of Remote Interface permit full remote control of the microhmmeter by an external
computer using either the RS-232 serial interface or the IEEE-488 parallel interface. The measured
value and instrument status can also be read by the computer.
The software commands are common to both interfaces and generally follow the IEEE488.2/SCPI
format. Either or both interfaces can be fitted to the instrument, however, only one of these can be in
use at any one time.
An additional interface can be fitted which permits more basic control of the microhmmeter by a
Programmable Logic Controller (PLC) or similar device. The Analogue Output feature is also included
on the PLC card.
Selection of the optional interfaces, together with additional features such as Baud Rate and addresses,
are set from the front panel menu.
All interface connectors are accessible on the rear panel.
The RS-232 interface requires Ver 2.0 software, or higher.
The IEEE-488 interface requires Ver 3.0 software, or higher.
The PLC interface requires Ver 4.0 software, or higher.
6.2.1 RS-232 SERIAL INTERFACE ( option MGR10-01 )
6.2.1.1 Connection
This conforms to specification ANSI/EIA/TIA-232-E-1991 "Interface Between Terminal Equipment
and Data Circuit-Terminating Equipment Employing Serial Binary Data Interchange." A list of pin
assignments is given in Appendix II.
Signal levels; MARK (logical "1"); -3V to -15V
SPACE (logical "0"); +3V to +15V
Data is transferred using the TXD and RXD lines.
Hardware handshaking for each character transfer uses the RTS and CTS lines.
RTS is an output from the microhmmeter which indicates its receiver status. When asserted (low) it
indicates that it is ready to receive another character. When negated (high) the microhmmeter receiver
buffer is full and cannot receive another character until the buffer is processed (if the PC sends one it
may be lost). As soon as space becomes available in the receiver buffer, RTS is re-asserted to allow the
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...