MGR10 A/B/C - Revision : I (for 7.0 software revision)
9
1.4 MGR10 and MGR10A technical specifications
1.4.1 Measurement Accuracy
Range Max.
Current Resolution
Accuracy at full rated current
Temp. Coefficient
30 k
100
A 1
±
(0.03%Rdg+0.02%FS)
50 ppm
3 k
1mA
100m
±
(0.03%Rdg+0.01%FS)
50 ppm
300
10mA
10m
±
(0.03%Rdg+0.01%FS)
50 ppm
30
100mA
1m
±
(0.03%Rdg+0.01%FS)
50 ppm
3
1A
100
±
(0.03%Rdg+0.01%FS)
50 ppm
200m
10A
10
±
(0.03%Rdg+0.01%FS)
50 ppm
30m
10A
1
±
(0.03%Rdg+0.01%FS)
50 ppm
3m
10A
100n
±
(0.03%Rdg+0.02%FS)
50 ppm
The accuracy is quoted at full rated current in SLOW measurement speed and the display count
is 30,000 plus sign and unit.
FS = Full Scale
1.4.2 Programmable current
The measurement current is adjustable on each resistance range as follows:
Range
Max Current
Min Current
Steps
30 k
100
A 10
A 1
A
3 k
1mA
100
A 10
A
300
10mA 1mA 100
A
30
100mA 10mA 1mA
3
1A 100mA
10mA
200m
10A 1A
100mA
30m
10A 1A
100mA
3m
10A 1A
100mA
The accuracy of current is
±
0.1% .
Open circuit voltage is 5 V +/-5%
1.4.3 Accuracy against current level
Range
100% Current
50% Current
10% Current
30 k
±
(0.03%Rdg+0.02%FS)
±
(0.04%Rdg+0.02%FS)
±
(0.05Rdg+0.02%FS)
3 k
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
300
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
30
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
3
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
200m
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
30m
±
(0.03%Rdg+0.01%FS)
±
(0.04Rdg+0.01%FS)
±
(0.05Rdg+0.01%FS)
3m
±
(0.03%Rdg+0.02%FS)
±
(0.04Rdg+0.02%FS)
±
(0.05Rdg+0.02%FS)
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...