MGR10 A/B/C - Revision : I (for 7.0 software revision)
29
Or press on
[Man]
to enter manually the compensation temperature .This mode request to the
operator to enter the ambiance temperature .
Press on
[Chg]
to modify the temperature value
Use the keypad to enter the new value in °C or °F
Press on [
OK
] to enter the new value
Press again on [
OK
] to go back to the previous page
Press on
[Cu]
to select compensation for copper material with a coefficient : 3980 ppm/
C
Or press on
[Al]
to select compensation for aluminum material with a coefficient : 4100 ppm/
C
Or press on
[User]
to select a coefficient which can be defined by the operator
Press on
[Chg]
to modify the coefficient value in ppm/°C
Use the keyboard to enter the new coefficient value
Press on [
OK
] to enter the new value
Press again on [
OK
] to go back to the previous page
Press again on [
OK
] to go back to the previous page
Ref :
Press on
[Ref]
to select the reference temperature value (standard and initial value set to 20°C)
Press on
[UP arrow]
or
[DOWN arrow]
keys to modify the temperature value
Press on [
OK
] to enter the new value
Press again on [
OK
] to go back to the previous page
On :
Press on
[On]
to activate the temperature compensation mode
Off :
Or press on
[Off]
to cancel the temperature compensation mode
Press on
[Quit]
to escape the menu
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...