MGR10 A/B/C - Revision : I (for 7.0 software revision)
78
*RST
Clears all pending operations, and resets the instrument to the default RESET condition as
described in the instrument user manual with two exceptions:
1) No change is made to the remote interface parameters (interface type,
baud rate and
IEEE address).
2) The trigger mode is set to SINGLE (INITiate:CONTinuous OFF).
For IEEE only, this also resets the Operation Complete Command State and Operation Complete
Query State (clears bit 0 of Standard Event Register).
The default conditions that are forced are:
Range:
30KOHM,
AUTO1
Current:
100%,
+I
Read Rate:
SLOW
Trigger Mode:
CONT:OFF (Single), STOP
Measurement Limits:
OFF, Alarm ON, Lower Limit = 0, Upper Limit = 30K
Filter:
OFF,
10
readings
Settling Algorithm:
OFF, 10 readings, 10 digits limit
Open Circuit Limit:
OFF
Temp Compensation:
OFF, MAN, +20
C, Cu, USER = 3980, Ref Temp = +20
C
Data Logger:
OFF, Count = 10, Datalog Memory unchanged.
READ? or FETCh?:
Set to FRESistance.
Talk Only mode: OFF
Analogue Output OFF
All other functions and parameters are left unchanged by *RST.
*TST?
Always returns ‘0’ to indicate self-test OK.
*WAI
This command is accepted by both interfaces, but ignored as all commands are executed
sequentially. It is provided only for compatibility with IEEE-488.2.
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...