MGR10 A/B/C - Revision : I (for 7.0 software revision)
71
needed for each measurement. If CONTinuous triggering is “on”, then repeated FETCh?
commands can be made to return the most recent measurement.
One of three optional functions can be included with this command:
FRESistance -
Returns the uncompensated resistance and is the default for the first FETCh?
TEMPerature -
Returns the temperature value of the external Pt100 probe, providing that
temperature compensation is “ON” and set to “EXT” mode. Otherwise, an
“execution error” is generated and the error value is returned.
TCOMpensate -
Returns the compensated resistance value at 20°C, providing that temperature
compensation is “ON”. Otherwise, an “execution error” is generated and the
error value is returned.
If no function is subsequently entered, then the function used in the previous FETCh? or READ?
command is used.
INITiate:CONTinuous <ON|OFF> or <1|0>
Setting to OFF means that any subsequent READ?, or INIT followed by FETCh?, will take one
reading and send it to the output buffer.
If set to ON, the microhmmeter will perform continuous measurements, which will be initiated
immediately, and FETCh? will send the most recent value to the output buffer. The measurements
can be stopped by setting CONTinuous to OFF.
(For the MGR10A model operating on battery power, continuous mode is not available, and
INITiate:CONTinuous ON will cause an “execution error” to be generated).
INITiate:CONTinuous?
Returns “0” (OFF) or “1” (ON).
READ[ FRESistance|TEMPerature|TCOMpensate]?
Triggers and returns a measurement. It is identical to INITiate, followed by FETCh?. See the
FETCh? command above, for use of additional functions.
This command is not available when the CONTinuous triggering mode is “on”, nor when
DATAlogging mode is “on”, in which cases, an “execution error” will be generated and the error
value returned.
ABORt
Stops the current measurement and clears the input and output buffers.
ABORt works for RS-232 only. If sent in IEEE-488 mode, it will generate a “command error”.
Содержание MGR10
Страница 1: ...SEFELEC REFERENCE PENT0060 Operating manual MGR10 English...
Страница 25: ...MGR10 A B C Revision I for 7 0 software revision 25...
Страница 68: ...MGR10 A B C Revision I for 7 0 software revision 68 autorange AUTO OFF AUTO1 AUTO2...
Страница 92: ...MGR10 A B C Revision I for 7 0 software revision 92 RS 232 Interface Commands SYSTem LOCal REMote...