TX180 Temperature Transmitter
15
The transmitter is indicating
FAILSAFE
, since no sensor is connected, and the analog output is
indicating greater than 100%, loop current at 23.00mA, which is the standard Failsafe report
condition. Please note that the display / keyboards can be plugged into the transmitter while the
transmitter is powered up. There is no need to disconnect power before plugging the LI-1 or LI-2 into
the TX180.
Press the key marked
NEXT
. The display starts to alternate asking if the user wishes to return to the
Operate Mode?
To activate the
NEXT
and
ENTER
; keys a slow, deliberate push of the key is required. This prevents
any casual, inadvertent activation of the transmitter into one of the configuration modes.
The answer would be “No”, therefore, press the
NEXT
key. This will enter you into the
DISPLAY
MODE
configuration menu. If you wish to answer a question “Yes”, press the
ENTER
key. A flow
chart summarizing the operation of the
DISPLAY MODE
appears at the end of this manual.
Note that when more than seven characters are required to describe a function, the display keeps
sequencing through two or more screens or may use common abbreviations. In this manual, the
sequencing of the display is indicated by placing the two or more parts of the message adjacently.
With some functions, the LI-2 display indicates a numeric value and unit of measurement on the top
line of the display in addition to the message on the lower display line.
4.2
D
ISPLAY
M
ODE
C
ONFIGURATION
The
DISPLAY MODE
will allow the user to do the following:
•
Select a Sensor Input (Select Input)
•
Select a desired temperature unit, such as ºF (Select Units)
•
Change the 4mA Lower Range Value (Change Zero)
•
Change the 20mA Full Scale Value (Change Full Scale)
•
Change the Sensor Fail Safe detection (Select Sensor Fail Safe)
•
Change the Fail Safe reporting (Select Fail Safe Report)
•
Trim the 4.0mA output current (Trim 4 MA)
•
Trim the 20.0mA output current (Trim 20 MA)
•
Trim the display value (Trim Display)
•
Change the Language of the display
Each of these functions is presented in sequence on the LCD display. If the indicated function need
not be performed, press Next, and the next function is displayed on the screen. To perform any
function press the Enter key. This will cause additional screens to be displayed which enable you to
perform the function. These are described in detail below and summarized on the LI-2 Two-Line
Display / Keyboard Flow Chart found in the rear of this booklet.
RETURN
TO
OPERATE
MODE?
Содержание TX180
Страница 1: ...Model TX180 Temperature Transmitter Installation and Operation Manual P N 974083 04 02 Rev B ECO 26721...
Страница 2: ......
Страница 26: ...26 Fig 7 Dynisco TX180 ITX190 Configuration Flowchart LI 2 Two Line Display Keypad...
Страница 40: ...40 Fig 8 Dynisco TX180 ITX190 Configuration Flowchart LI 2 One Line Display Keypad...
Страница 50: ...NOTES...
Страница 52: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...