TX180 Temperature Transmitter
11
2.3.2.3 T
HREE
-W
IRE
RTD I
NPUT
Apply the common legs from the RTD (generally the same color RTD leads) to terminals 1 and 2.
Apply the other signal lead to terminal 3. Terminals 3 and 4 must be jumpered together for proper
operation and to prevent any build-up of electrostatic charge on these terminals which could affect
the transmitter readings.
2.3.2.4 F
OUR
-W
IRE
RTD I
NPUT
Apply one set of the common legs from the RTD (generally the same color RTD leads) to terminals 1
and 2. Apply the other signal lead pair to terminals 3 and 4. No jumpers are necessary for a 4 wire
RTD input.
3.
T
RANSMITTER
O
PERATION
3.1
I
N
A
HURRY
?
When in a hurry, this short set of instructions and references will help get the transmitter running.
3.1.1 F
ACTORY
C
ONFIGURATION
Input
=
Type J Thermocouple
Output
=
Analog
4.00mA
=
40
°
F
20.00mA
=
200
°
F
Sensor Fail-safe
=
23.00mA (High)
On special request the factory will set the transmitter to any desired configuration. Special
configurations are identified on a tag attached to the unit.
3.1.2 O
PERATION
W
ITHOUT
A
D
ISPLAY
If the unit was ordered with the standard factory configuration, the sensor required is a Type J
thermocouple. The packing slip and a tag on the unit will indicate if the unit was set up to any other
customer requested special configuration. If there is a need to change the configuration of the
transmitter, or to re-range it, use either the LI-1 or LI-2 Display and refer to the procedures described
in SECTIONS 4 (for LI-2), 5 (for LI-1).
NOTE:
Even when “In a Hurry”, the use of an appropriate power supply is important. A 24V DC
supply having a current handling capacity of at least 0.1A is commonly used. Always use a
DC (direct current) supply, or suitable size battery.
NEVER CONNECT THE TRANSMITTER
DIRECTLY TO 115VAC.
With the power supply off, connect the + side of the power supply to the + terminal of the
Содержание TX180
Страница 1: ...Model TX180 Temperature Transmitter Installation and Operation Manual P N 974083 04 02 Rev B ECO 26721...
Страница 2: ......
Страница 26: ...26 Fig 7 Dynisco TX180 ITX190 Configuration Flowchart LI 2 Two Line Display Keypad...
Страница 40: ...40 Fig 8 Dynisco TX180 ITX190 Configuration Flowchart LI 2 One Line Display Keypad...
Страница 50: ...NOTES...
Страница 52: ...DYNISCO INSTRUMENTS 38 FORGE PARKWAY FRANKLIN MA 02038 ATTN MARKETING DEPT Place Stamp Here...