Dynatron
®
150
plus
Calibration Procedure
42
Te soundhead type is displayed for about
one second. Next, the display shows the
stored frequency values for the
soundhead that is being calibrated. The
output power display is set to a value of
zero. The LED for 1 MHz should be
illuminated indicating the soundhead is
now ready to be calibrated for this
frequency. If it is not, press the toggle
key to select 1 MHz now.
3. Press
STOP and the TIME UP ARROW
simultaneously to enter the Temperature
Mode. The temperature of the soundhead
is displayed in the time display window.
The soundhead must be at a temperature
between 72 and 75 degrees. If the
soundhead is not within this range, warm the water and/or the head before
continuing. Press STOP to get out of the temperature mode. At this point there is
zero output from the device. Zero the scale on your Ultrasound power meter now.
When the scale is at zero on your power meter, press the START key to begin.
4. Press
the
UP/DOWN ARROW keys located next to the Power display window on
the Dynatron 150plus until the power meter shows 1.0 W/cm
2
.
5. When the watts reading on the meter is at 1.0 W/cm
2
for the soundhead and the
reading is stable, press the DUTY CYCLE toggle key (this is the key that selects
10%, 20%, 50% or Continuous).
6. The Dynatron 150plus performs the calculation internally for the “Z” (impedance
value) and the coupling value. These values are automatically entered and stored in
the Head Parameters for the device (this operation is automatic and invisible to the
user). Record these values on the calibration sheet for future reference.
7. Press the frequency toggle key to select 2 MHz. The 2 MHz LED will light
indicating the soundhead is now ready to be calibrated at the frequency. Repeat
steps 3 through 6 above.
8. Press the frequency toggle key to select 3 MHz. The 3 MHz LED will light
indicating the soundhead is now ready to be calibrated at the frequency. Repeat
steps 3 through 6 above.
9. If you wish to cancel the calibration procedure without storing any new
calculations, press the PAUSE key.
10. This completes the calibration of one soundhead. For devices that use more than one
soundhead, you must calibrate each of those soundheads individually.
Soundhead Types Table
Smart Heads:
Time Display = HS
Power
Display:
2 cm2 = 2
5 cm2 = 5
10
cm2
=
10
Older Sound Heads:
Time Display only:
1 cm2 = H 1
2 cm2 = H 2
5 cm2 = H 5
10 cm2 = H10
Содержание 150plus
Страница 1: ...SERVICE MANUAL Dynatron 150plus ...
Страница 6: ...Dynatron 150plus vi ...
Страница 20: ...Dynatron 150plus 14 ...
Страница 24: ...Dynatron 150plus 18 ...
Страница 32: ...Dynatron 150plus 26 ...
Страница 42: ...Dynatron 150plus Beam Profile 36 2 cm2 Head Near Field 1 MHz 3 MHz ...
Страница 46: ...Dynatron 150plus 40 ...
Страница 50: ...Dynatron 150plus 44 ...
Страница 54: ...Dynatron 150plus 48 ...
Страница 59: ...53 ...
Страница 60: ...54 ...
Страница 61: ...55 ...
Страница 62: ...56 ...
Страница 63: ...57 ...
Страница 64: ...58 ...
Страница 65: ...59 ...