Dynatron
®
150
plus
Contraindication, Warnings, Precautions
28
POWER SHOULD BE REDUCED IF PATIENT COMPLAINS OF
PERIOSTEAL BONE PAIN (BONE ACHE)
Precautions
The Dynatron 150plus must be used cautiously in the presence of any of the
following conditions:
• When there is a tendency to hemorrhage following acute trauma or fracture.
• Acute Bursitis: Do not use in continuous duty cycle mode.
Warnings
• Do not use in general area where high-powered, high-frequency transmitting units
are being operated. Short wave diathermy should not be turned on or used at the
same time as the Dynatron 150plus.
• Do not use the same power outlet or line with a whirlpool and certain traction
machines.
• In areas which are carpeted and static electricity is present, it may be necessary to
use a conductive mat to remove any static charge from the operator.
• Use a surge suppresser if power problems are encountered.
• Avoid unnecessary exposure to Ultrasound (patient and therapist)
NOTE: For further technical information on therapeutic Ultrasound and its
physiological effects, please consult published medical literature.
Usage Cautions - Combination Treatments
When using a Stim device in conjunction with the Dynatron 150plus to output Stim
through the soundhead, observe all contraindications, warnings, precautions, and
usage cautions provided by the manufacturer for that Stim therapy along with all
contraindications, warnings, precautions, and usage cautions for Ultrasound therapy
provided in this manual.
Safety Features of the Dynatron 150plus
• The device senses coupling and soundhead temperature to help ensure proper
delivery of the ultrasonic therapy and to protect the soundhead crystal.
• The device provides complete isolation of patient connection from AC power
line.
Содержание 150plus
Страница 1: ...SERVICE MANUAL Dynatron 150plus ...
Страница 6: ...Dynatron 150plus vi ...
Страница 20: ...Dynatron 150plus 14 ...
Страница 24: ...Dynatron 150plus 18 ...
Страница 32: ...Dynatron 150plus 26 ...
Страница 42: ...Dynatron 150plus Beam Profile 36 2 cm2 Head Near Field 1 MHz 3 MHz ...
Страница 46: ...Dynatron 150plus 40 ...
Страница 50: ...Dynatron 150plus 44 ...
Страница 54: ...Dynatron 150plus 48 ...
Страница 59: ...53 ...
Страница 60: ...54 ...
Страница 61: ...55 ...
Страница 62: ...56 ...
Страница 63: ...57 ...
Страница 64: ...58 ...
Страница 65: ...59 ...