8 | Maintenance
D-R 290
173
Se
qu.
no.
Critical fault
Cause;
(●) available actions
by whom
[194] Hardware fault
The electronic hardware of the device has a serious
internal fault.
●
Exchange the electronics (No. 22)
Service technician
[195] Zero point reflector positioning
error
The zero point reflector could not find the light barri
er.
●
Check the operation of the stepper motors
Operating
personnel
●
Possibly exchange the light barrier
Service technician
[196] Turntable position error
The turntable could not find the light barrier.
●
Check the operation of the stepper motors
Operating
personnel
●
Possibly exchange the light barrier
Service technician
[197] LED current too low
No current through the LED, or very little current
●
Check the plug / contact of the LED connecting
cable
Operating
personnel
●
LED possibly defective: Exchange the LED
Service technician
[198] LED current too high
Current through the LED is too high
●
LED possibly defective: Exchange the LED
●
Error in activating the LED, electronics (no. 22)
Service technician
[199] Fault fail-safe shutter 1
(measuring head)
Switch contact fault at the fail-safe shutter protect
ing the measuring head
●
Check the connections for the fail-safe shutter
protecting the measuring head, and check the
control device
Specialised
personnel
[200] Fault fail-safe shutter 2
(reflector)
Switch contact fault at the fail-safe shutter protect
ing the reflector
●
Check the connections for the fail-safe shutter
protecting the measuring head, and check the
control device
Specialised
personnel
[201] Fail-safe shutter 1 blocked
(measuring head)
Fail-safe shutter protecting the measuring head
does not open or close within 3 min.
●
If the supercaps for starting the device were
completely discharged, restart the device
●
Check the operation of the fail-safe shutter
●
Check the connections for the fail-safe shutter
protecting the measuring head, and check the
control device
Specialised
personnel
[202] Fail-safe shutter 2 blocked
(reflector)
Fail-safe shutter protecting the reflector does not
open or close within 3 min.
●
If the supercaps for starting the device were
completely discharged, restart the device
●
Check the operation of the fail-safe shutter
●
Check the connections for the fail-safe shutter
protecting the measuring head, and check the
control device
Specialised
personnel
Table 8.5: List of critical faults for the D‑R 290
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...