8 | Maintenance
170
D-R 290
Se
qu.
no.
Warning
Cause;
(●) available actions
by whom
[090] Temperature measurement of
the stepper motor is defective
The temperature sensor of the stepper motor is de
fective
●
No effect on the operation of the device, but:
Protective function / automatic switch-off of the
stepper motors if the temperature is too high or
too low is no longer active
Service technician
[091] Warning of contamination
The device contamination is higher than the config
urable warning threshold
●
Clean the device plate, zero point reflector and
reflector unit
Specialised person
nel
[092] Purge air flow 1 (measuring
head) too low
Purge air flow is below the warning threshold. Pos
sible causes: Filter clogged, purge air hose split,
parameterisation of the purge air inlet defective.
●
Measurement function still operates, but a rapid
increase in contamination can be expected.
Check the purge air flow / restore the flow or re
move the device from the flue stack.
Specialised person
nel
[093] Purge air flow 2 (reflector) too
low
Purge air flow is below the warning threshold. Pos
sible causes: Filter clogged, purge air hose split,
parameterisation of the purge air inlet defective.
●
Measurement function still operates, but a rapid
increase in contamination can be expected.
Check the purge air flow / restore the flow or re
move the device from the flue stack.
Specialised person
nel
[094] Purge air flow 1 (measuring
head) too high
Purge air flow above the warning threshold. If the
purge air flow is too high, depending on the flue
stack temperature and flue gas velocity there may
be an effect on the measured values.
●
Check the effect of the purge air on the meas
ured values
Specialised person
nel
[095] Purge air flow 2 (reflector) too
high
Purge air flow above the warning threshold. If the
purge air flow is too high, depending on the flue
stack temperature and flue gas velocity there may
be an effect on the measured values.
●
Check the effect of the purge air on the meas
ured values
Specialised person
nel
[096] LED heating defective
No current flowing through the LED heating
●
Measurement function still operates, but at very
low temperatures (< -30°C/-22°F) a protective
shut-down of the LED may occur
Service technician
[097] Stepper motor heating defec
tive
No current flowing through the stepper motor heat
ing
●
Measurement function still operates, but at very
low temperatures (< -40°C/-40°F) a protective
shut-down of the stepper motor may occur
Service technician
[098] Disc heater defective
No current flowing through the disc heater
●
Measurement function still operates, but if tem
peratures fluctuate greatly and at high atmos
pheric humidity, condensation may occur on the
device disc
Service technician
Table 8.3: List of warnings for the D‑R 290
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...