7 | Checking / setting parameters with D‑ESI 100
D-R 290
129
Type
Unit
Description
Opacity
% OP
Opacity as specified in section
. Calcula
ted using the current stack correction factor that was set
(see
)
Optical density
OD
Optical density as specified in section
Opacity (SP)
%OP SP
Optical density as specified in section
with a fixed flue stack correction factor of 1.
Opacity (DP)
%OP DP
Optical density as specified in section
with a fixed flue stack correction factor of 2.
Extinction coefficient
1/km
Extinction coefficient
k as specified in section
tion of the dust concentration [
.
For correct calculation of the extinction coefficient, the
measurement path length must be set (see
Visual range
km
Koschmieder visual range. For correct calculation of the
visual range, the measurement path length must be set
(see
)
Table 7.4: Available type codes for the D-R 290
Example: Settings for the measured value correction
If necessary the real measurement value that was determined can be modified to suit
the physical characteristics of the duct in which the measurement is being performed,
and/or to suit the purpose to which the signal is to be put (depending on the plant or
evaluation process).
A measured value correction is performed irrespective of any gravimetric calibration
parameter setting based on a gravimetric calibration.
If a measured value correction is performed, this takes effect in addition to the settings
for the gravimetric calibration.
A measured value correction is performed using the variables "Offset a0", "Gradi
ent a1", "Gradient a2" and "Gradient a3".
Setting path:
Sitemap > »Device (e.g. D‑R 290 ) > >
Common Parameters > Setting
variables > [offset a0/gradient a1/
gradient a2/gradient a3] >
Adjusting the variable for the measured value correction
Left click on "Device" in the sitemap.
Left click on
for the desired device in the device list.
Under
Common parameters
in the list field, after "
Select group
", select "
Set
ting variables"
.
7.5.3
Step 1
Step 2
Step 3
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...