8 | Maintenance
D-R 290
171
Se
qu.
no.
Simple fault
Cause;
(●) available actions
by whom
[128] System settings invalid
The device settings held in the EEPROM are inva
lid. The device may be operating with invalid pa
rameters.
●
Check the device settings and correct them if
necessary. After the correction, save the
changes!
Specialised
personnel
●
If the fault persists, the electronics are defective. Service technician
[129] Invalid firmware
The device firmware may not be operating correct
ly.
●
Check the operation of the device and any other
messages, reboot the device.
Service technician
●
If the fault persists, the electronics are defective.
[130] Device not initialised
The device is not correctly parameterised, or the
parameterisation has been lost due to a memory
defect.
Service technician
[131] Receiver overrun
The signal intensity is too high.
●
Check the beam inlet, comply with the minimum
distance between the reflectors
Specialised
personnel
[132] Receiver overrun
Electronics fault.
Service technician
[133] Receiver overrun
Electronics fault.
Service technician
[134] Receiver overrun
Electronics fault.
Service technician
[135] Contamination error
The contamination is greater than 10%.
●
Clean the device plate, zero point reflector and
reflector unit
Specialised
personnel
[136] LED undertemperature shut-
down
The LED was shut down because the LED temper
ature was too low.
●
If the device is started up at a very low tempera
ture, the warming up of the device components
can take up to 30 min.
●
Check the device temperature and
the current in the heaters
Specialised
personnel
[137] LED overtemperature shut-
down
The LED was shut down because the LED temper
ature was too high.
●
Provide cooling.
Specialised
personnel
[138] Stepper motor undertempera
ture shut-down
The stepper motor was shut down because the
stepper motor temperature was too low.
●
If the device is started up at a very low tempera
ture, the warming up of the device components
can take up to 30 min.
●
Check the device temperature and the current in
the heaters
Specialised
personnel
[139] Stepper motor overtempera
ture shut-down
The stepper motor was shut down because the
stepper motor temperature was too high.
●
Provide cooling.
Specialised
personnel
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...