7 | Checking / setting parameters with D‑ESI 100
D-R 290
127
Setting path:
Sitemap > »Device (e.g. D‑R 290 ) > >
Save Parameters
For security reasons, parameters that are changed (e.g. specific parameters such as
data for gravimetric calibration) are initially written only to the working memory of the
device. This means these data are currently used or referenced, but are no longer
available after the device has been restarted (i.e. the device then reverts to using the
"old" parameters).
Clicking on the "Save parameters" button causes the data to be transferred to the
read-only memory of the device. This means the parameters are also retained and
continue to be used even the device is restarted.
For the new setting to be used permanently, left click on the "
Save parameters
"
button. (The new setting is not used permanently until it has been saved.)
Type code
The "Type code" selects the output variable for the respective channel.
The variables setting for the "Type code" permits a selection based on a code table.
A DURAG device (measuring head) generally has 4 measured value output channels.
In each channel the output of the relevant measured value (see diagram below) is par
ameterised by clicking on the appropriate radio button.
Close ≘ (abort and) close window
Send ≘ load (temporarily)
Table 7.3: Measured value output channel
» Device
» Options
» Data log
» Login
Step 6
Save parameters
7.5.2
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...