B-12
TR2510 Current Probe
TR2510 Current Probe
Guidelines
To achieve the rated accuracies, follow these guidelines:
• The conductor must be at a right angle to the probe.
• The conductor must be centered in the probe core.
• The jaw contact surfaces must be clean and properly aligned.
Specifications
Range: 0.1 to 10 Arms
Accuracy (at 60 Hz): 0.1 to 10A, ±1% reading ±0.01A
Working voltage: 600 V max.
Frequency range: 40 to 3000 Hz
Output signal: 150 mV/A
Maximum conductor size: 0.78" (20 mm)
Phase shift (at 60 Hz):
1 to 5A less than 1º
5 to 10A less than 1.5º
Operating temperature: 14 to 131 ºF (-10 to 55 ºC)
Storage temperature: -40 to 158 ºF (-40 to 70 ºC)
Dimensions: 5.47 x 2.00 x 1.18” (139 x 51 x 30 mm)
Weight: 6.5 oz. (180 g)
NOTE
Current probe TR2510 can be used interchangeably with TR2510A.
www.
.com
1.800.561.8187
Содержание PowerGuide 4400
Страница 13: ...xiii Dranetz PowerGuide 4400 www com information itm com 1 800 561 8187 ...
Страница 14: ...xiv www com information itm com 1 800 561 8187 ...
Страница 78: ...4 16 www com information itm com 1 800 561 8187 ...
Страница 136: ...6 22 www com information itm com 1 800 561 8187 ...
Страница 158: ...7 22 www com information itm com 1 800 561 8187 ...
Страница 196: ...9 16 www com information itm com 1 800 561 8187 ...
Страница 206: ...A 10 www com information itm com 1 800 561 8187 ...
Страница 226: ...C 8 www com information itm com 1 800 561 8187 ...
Страница 228: ...D 2 www com information itm com 1 800 561 8187 ...
Страница 252: ...F 4 www com information itm com 1 800 561 8187 ...
Страница 262: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9 www com information itm com 1 800 561 8187 ...
Страница 272: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19 www com information itm com 1 800 561 8187 ...