C-3
APPENDIX C/ Battery Specifications and Replacement Procedure
Battery Safety Precautions
WARNING
DO NOT intentionally short circuit the battery pack. The batteries are capable of
providing hazardous output currents if short circuited. The 4400 is equipped with
an internal battery charger circuit. Do not attempt to charge the batteries with an
external charger other than the Dranetz battery charger, since improper
charging could cause battery explosion.
ADVERTENCIA
NO ponga intencionalmente la bateria en cortocircuito. Las baterias son capaces
de proporcionar corrientes de salida peligrosas si est·n en cortocircuito. La 4400
esta equipada con un circuito interno cargador de baterÌas. No intente cargar las
baterÌas con un cargador externo que no sea el cargador de baterias Dranetz
,
puesto que la carga indebida podrÌa hacer que explote la bateria.
AVERTISSEMENT
NE PAS court-circuiter délibérément le bloc-batterie. Lors díun court-circuit, les
batteries risquent díémettre des courants effectifs dangereux. 4400 posséde un
circuit de chargeur de batterie intégré. Ne pas tenter de charger les batteries au
moyen díun chargeur externe autre que le chargeur de batterie Dranetz, car
un rechargement fautif pourrait entrainer l’explosion de la batterie.
WARNUNG
Die Batterien dürfen NICHT kurzgeschlossen werden. Im Falle eines
Kurzschlusses konnen die Batterien lebensgefährliche Ausgangsstrome leiten.
4400 ist mit einem internen Batterieladegerät ausgestattet. Die Batterien sollten
nur mit dem Ladegerät von Dranetz geladen werden. Die Verwendung eines
externen Ladegeräts kann zu einer Explosion der Batterien führen.
Safety
precautions
Observe the following precautions when inspecting or replacing the battery pack:
• Do not attempt to replace individual batteries of the pack or substitute other battery
types.
• Do not dispose of battery in fire.
• Dispose of a used battery promptly in accordance with local Environmental
Protection Agency (EPA) regulations.
• Visually inspect the battery pack for corrosion.
The batteries have a safety pressure vent to prevent excessive gas build-up and
corrosion indicates that venting has occurred. Possible causes of venting are: a
defective charger, excessive temperature, excessive discharge rate, or a defective
cell.
If corrosion is excessive, the battery pack may require replacement (contact Dranetz
Customer Service Department).
www.
.com
1.800.561.8187
Содержание PowerGuide 4400
Страница 13: ...xiii Dranetz PowerGuide 4400 www com information itm com 1 800 561 8187 ...
Страница 14: ...xiv www com information itm com 1 800 561 8187 ...
Страница 78: ...4 16 www com information itm com 1 800 561 8187 ...
Страница 136: ...6 22 www com information itm com 1 800 561 8187 ...
Страница 158: ...7 22 www com information itm com 1 800 561 8187 ...
Страница 196: ...9 16 www com information itm com 1 800 561 8187 ...
Страница 206: ...A 10 www com information itm com 1 800 561 8187 ...
Страница 226: ...C 8 www com information itm com 1 800 561 8187 ...
Страница 228: ...D 2 www com information itm com 1 800 561 8187 ...
Страница 252: ...F 4 www com information itm com 1 800 561 8187 ...
Страница 262: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9 www com information itm com 1 800 561 8187 ...
Страница 272: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19 www com information itm com 1 800 561 8187 ...