9-8
External Communication Interface, continued
Connection
setup via
Modem
(continued)
Back at the 4400, the last step in configuring the modem communications is to modify
the modem initialization string to match the modems and chipsets being used. The goal
is to configure for the following:
• Restore factory settings.
• Set auto-answer to answer on one or two rings (more than two rings are allowed but
at some point the number of rings may start to conflict with NodeLink
®
timeouts).
• Turn hardware flow control off since you are using a "two-wire" connection (fiber
optic port).
• Set a modem inactivity timeout of 60 seconds. The modem will hang up after one
minute of continuous inactivity on the line.
• Select the minimum allowable baud rate.
Follow the steps below to change the modem initialization string. See Chapter 4
Instrument Settings - Communications for the 4400 screen interface.
1. From the 4400
Communications Settings
screen used to configure modem
connection speed, click on the
Change
button to the right of the string field display.
2. Use the virtual keypad that appears on the screen to change and enter the new string
command. The modem initialization string commands for the Conexant (Rockwell)
chipset and the Agere (Lucent) chipset are discussed in the next section.
3. When done, click on
OK
to accept changes in modem communications settings.
Modem
initialization
string command
The following is the AT initialization string for a modem using the newer Conexant
(Rockwell) chipset:
AT&F;S0=2;+IFC=0,0;S30=6
Where:
AT is the modem attention command
&F instructs the modem to restore factory configuration
S0=2 turns auto-answer on and sets the number of rings to answer to two.
+IFC=0,0 turns flow control off.
S30=6 sets a 60 second inactivity modem hang up timeout.
; is a command delimiter.
NOTE: It may be necessary to adjust the modem connection speeds to specify a
minimum baud rate. This is possible using the +MS modem command. Check the
modem manual for use of the +MS command. There are several varieties of the
command. Typically, a modem modulation scheme and minimum and maximum
transmit and receive speeds can be specified. Using Hyperterminal, it is possible to get
a list of the possible command ranges and syntax by typing AT+MS=?.
Continued on next page
www.
.com
1.800.561.8187
Содержание PowerGuide 4400
Страница 13: ...xiii Dranetz PowerGuide 4400 www com information itm com 1 800 561 8187 ...
Страница 14: ...xiv www com information itm com 1 800 561 8187 ...
Страница 78: ...4 16 www com information itm com 1 800 561 8187 ...
Страница 136: ...6 22 www com information itm com 1 800 561 8187 ...
Страница 158: ...7 22 www com information itm com 1 800 561 8187 ...
Страница 196: ...9 16 www com information itm com 1 800 561 8187 ...
Страница 206: ...A 10 www com information itm com 1 800 561 8187 ...
Страница 226: ...C 8 www com information itm com 1 800 561 8187 ...
Страница 228: ...D 2 www com information itm com 1 800 561 8187 ...
Страница 252: ...F 4 www com information itm com 1 800 561 8187 ...
Страница 262: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9 www com information itm com 1 800 561 8187 ...
Страница 272: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19 www com information itm com 1 800 561 8187 ...