B-11
APPENDIX B/ Technical Specifications
TR2500 Current Probe
Guidelines
To achieve the rated accuracies, follow these guidelines:
• The conductor must be at a right angle to the probe.
• The conductor must be centered in the probe core.
• The jaw contact surfaces must be clean and properly aligned.
Specifications
Range: 1 to 500 Arms
Accuracy:
48 to 400 Hz: ±1.5% reading ±0.6A
400 to 3000 Hz: ±3% reading ±0.8A
Working voltage: 600 V max.
Frequency range: 48 to 3000 Hz
Output signal: 3 mV/A
Maximum conductor size: 1.18" (30 mm)
Maximum bus bar size: 2.5 x 0.20" (63 x 5 mm)
Phase shift (45 to 1000 Hz): Less than 4º
Operating temperature: 14 to 122 ºF (-10 to 50 ºC)
Storage temperature: -40 to 158 ºF (-40 to 70 ºC)
Dimensions: 2.60 x 7.68 x 1.34” (66 x 195 x 34 mm)
Weight: 14.82 oz. (420 g)
NOTE
Current probe TR2500 can be used interchangeably with TR2500A.
www.
.com
1.800.561.8187
Содержание PowerGuide 4400
Страница 13: ...xiii Dranetz PowerGuide 4400 www com information itm com 1 800 561 8187 ...
Страница 14: ...xiv www com information itm com 1 800 561 8187 ...
Страница 78: ...4 16 www com information itm com 1 800 561 8187 ...
Страница 136: ...6 22 www com information itm com 1 800 561 8187 ...
Страница 158: ...7 22 www com information itm com 1 800 561 8187 ...
Страница 196: ...9 16 www com information itm com 1 800 561 8187 ...
Страница 206: ...A 10 www com information itm com 1 800 561 8187 ...
Страница 226: ...C 8 www com information itm com 1 800 561 8187 ...
Страница 228: ...D 2 www com information itm com 1 800 561 8187 ...
Страница 252: ...F 4 www com information itm com 1 800 561 8187 ...
Страница 262: ...MARK201 LEVEL 200 Scope Mode MARK202 MARK205 G 9 www com information itm com 1 800 561 8187 ...
Страница 272: ...MARK112b LEVEL 112_701 Advanced Options EN50160 Options MARK112c G 19 www com information itm com 1 800 561 8187 ...