5. EIT-Terminology
Baseline:
Sets the reference for the next breath or sequence of breaths (changes in
impedance). The virtual baseline EIT image in the different views is black.
ΔEELI = Changes of EELI:
Changes in end-expiratory lung impedance in the EIT sensitivity region.
Absolute impedance measurements cannot be directly related to end-
expiratory lung impedance. Nevertheless, ∆EELI is strongly correlated to
changes in the end-expiratory lung volume within the EIT sensitivity region.
The PulmoVista
®
500 parameter ∆EELI expresses regional end-expiratory
lung impedance variations compared to the global tidal variation of Cursor
Ref.
EELv = end-expiratory lung volume:
EELV is often used interchangeably with the term FRC. However, mechanically
ventilated patients exhale against PEEP and not against ambient pressure.
Therefore, physicians use the term EELV instead of FRC.
Globale Impedance Curve
D
-11958-2017