Different views and their special properties
| 21
Measurement in 15-electrode mode
2. different views and
their special properties
Different views are available for displaying and evaluating the EIT images and
image information.
–
Main view
–
Fullscreen view
–
End-inspiratory trend view
–
End-expiratory trend (∆EELI) view
–
Diagnostics view
2.1. MaIn and FULLSCrEEn vIEw
After starting the measurement, you are in the
main view
. This view provides an
overview of the distribution of the tidal volume in the transverse EIT sensitivity
region. In the dynamic image, the volume distribution is displayed in real time at
up to 50 fps, inspiratory and expiratory, so inhomogeneities of the lungs can be
quickly identified. The
Tidal Image captures the volume distribution of the last
detected breath. To assess regional ventilation during spontaneous breathing
D