30 |
2.3.2. rESULTS OF ThE anaLYSIS
The results of the PEEP trial analysis support the selection and
personalization of PEEP. Regional compliance images show the influence
of PEEP on lung mechanics. The ›Customized analysis‹ can be used to
evaluate various therapeutic interventions.
At constant driving pressures an increase/decrease in ventilation can be
interpreted as an increase/decrease of regional compliance (C).
1
2
3
4
5
6
1
Global impedance waveform and sections. If available, trend parameters are
displayed: PEEP in green and EIP (or PIP) in pink.
2
Tidal Image for every displayed section.
3
Compliance loss image for every displayed section.
4
Areas in which a regional ventilation delay was detected and corresponding
RVD parameter, displayed in yellow.
5
Diagram with the numeric values of all sections depending on the settings and
analysis type.
6
Create report: An image and a text file are created containing the entered
parameters, the event list, all available analysis parameters and MEDIBUS values,
and a screenshot of the analysis. (Only active if a USB mass storage device is
connected). This report is meant to be used for documentation purposes.
D
-30761-2017