Prepare the device for the clinical use
| 11
Patient cable
Right port
Electrode belt
Left port
D
-6058-2014
1.2.3. Prepare the patient
Before the electrode belt is put on, check the potential necessary preparations
to position it well. For example, wound dressings, drainages, ECG electrodes,
catheters, strong hair growth or surgical and disease-related problem areas
may impair the placement of the belt or electrodes. If possible, the contact
area should be exposed to improve skin contact. However, if one electrode has
insufficient skin contact, the EIT measurement can also be started in the so-called
15-electrode mode. Further information can be found in the chapter
›
Checking
signal quality
‹.
D
-11932-2017
D
-11931-2017
When attaching the electrode cable avoid too tight bending radii, especially at
the transition points between electrodes 4/5 and 12/13 (port left/right).