1.2.4. Prepare the electrode belt
To be ready for use as quickly as possible, it is advisable to moisten the
electrodes of the electrode belt with a small amount of liquid (e.g. sterile saline
solution, water or electrode gel), in particular with dry skin.
1.2.5. Positioning of the electrode belt on the thorax
For orientation and easier alignment there is a noticeable silicone marking on the
electrode belt. This
mid position marker is located between the 8
th
and 9
th
electrode and should be positioned on the spine between the 4
th
and 6
th
ICS.
1.2.6. General procedure when attaching the electrode belt
Apply the belt to the supine patient by
–
lifting the patient’s upper body or
–
rolling the patient to the side.
D
-2284-2010
Mid position marker
If the patient condition allows it, it may
also be advisable to bring the bed in an
elevated position. This makes it easier to put
on the belt. The EIT measurement can be
performed regardless of the position, e.g.
in seated, supine, prone or lateral position.
45° to 80°
D
-11942-2017
Содержание PulmoVista 500