38 |
3.5. COUPLInG OF a drÄGEr vEnTILaTOr
wITh PULMOvISTa
®
500
PulmoVista
®
500 allows a synchronous import of EIT information and the current
ventilation settings, waveforms, and trends. Data is imported via the Medibus /
Medibus.X interface. The necessary settings can be found in the menu under
›Screen layout / Data Import‹. For data transmission, the following settings must
be made at the ventilator:
Baudrate:
19200
Parity:
non
Stopbit:
1
The data cable is connected to the COM1 port of PulmoVista
®
500.
After successful pairing, the ventilator name is displayed in the menu.
D
-30756-2017
Data import menu with configuration of curves and
parameters for trend views