Safety Station Technical Service Manual
5
10/17/2011
GD-99009 Rev. B
Tech n i cal D escri pt i o n
S p e c if ic a t ion s
The TPSS specifications are presented in Table 1 below. TPSS external dimensions and features are shown in
Figures 1 and 2 on the following pages.
TABLE 1
Item
Transposal Safety Station (#TP-SS100)
Certifications
Certified and Listed as a Class 1 Medical Device as
authorized under the Medical Device Standard IEC
60601-1 and IEC 60601-1-2 for EMC.
Certified to CAN/CSA Standard C22.2 No. 601.1.
IP Certification Classification
IP 60
IAPMO Listed
IGC-238-2007
Patents
5,807,359; 5,931,822; 6,263,887; 6,368,310;
6,588,436; 6,673,055; 6,776,175; 6,796,317
Installation
Fixed
Size (inch (cm))
36(91)W x 28(71)D x 48(122)H
Weight (lbs (kg))
260(118)
Utilities
Power
120VAC, 60Hz, 15amp (See Replacement Items for
fuse specifications).
Water
3/4" Hot and Cold, 6.0gpm(22.7lpm), 55psi
Hot Water Temperature 110 – 120
o
F (43 - 49
o
C)
Drain
2"
Space
37(94)W x 40(102) D x 66(168)H
Additional Cooling
None.
Flammable Rating
This equipment is not suitable for use in the presence
of flammable anesthetic mixture with air or with
oxygen or with nitrous oxide.
Operation
Mode
Continuous
Cycle
128 to 210 seconds
Water Use
7.0(26.5) to 11.2(42.4) gallons(liters)/cycle
Enzyme Use
0.3(8.9) to 0.7(20.7) ounces(milliliters)/cycle
Bleach Use
3.4(100.6) ounces(milliliters)/cycle
Method of Sterilization
N/A