Safety Station Technical Service Manual
13
10/17/2011
GD-99009 Rev. B
Performing Verify Disinfection Cycle:
To verify that the unit is disinfecting, place the unit into the check for disinfection mode by using the following
steps:
1. Place and lock two canisters into the rings.
2. Close the lid.
3. Press the “Wash Button” then depress the “Clean Sink” button twice.
4. The main display will indicate that the unit has gone into the disinfection test mode.
5. When the unit reaches the point in the cycle where the disinfection solution is on the canisters, unit will
stop and alarm.
6. Press the “open” button to open the lid.
7. Use the test strips, included in the in service tub, to wipe the inside and outside of each canister.
8. If the test strip turns purple the canisters are being disinfected.
9. If the test strips do not turn a different color, call DMS at 1-888-466-6633 and ask for service.
10. 10. After testing is complete, close the lid to finish the processing cycle.
Clean Sink Cycle
To run a clean sink cycle and disinfect the unit’s basin, verify that there are no canisters inserted into the unit’s
rings. Close the lid and press the “clean sink” button on the units control panel. The lid will open automatically
at the end of the clean sink cycle.
Single Use Lids
The TP-DL2800 lid is designed to be used for a single patient. Failure to change the lid after each patient may
result in the following:
1.
Loss of available suction. Contaminated and/or wet filters will affect vacuum flow.
2.
Cross contamination. Improper usage of single use items poses a health risk to patients and health
care providers.
Exterior Cleaning and Disinfection
The exterior of the unit can be decontaminated by using an approved hospital grade disinfectant. It is
recommended that Personal Protective Equipment (PPE) be worn at all times when doing so.
Storage Between Uses
The DMS TPSS requires no special shutdown or storage procedures between uses. It may remain plugged in and
connected to water supplies.