Safety Station Technical Service Manual
15
10/17/2011
GD-99009 Rev. B
T r o ub lesh o o tin g Gu id e
Problem: Inadequate Cleaning of Canister
Possible Solution:
1. Check drain. If the drain is slow or clogged, call maintenance.
2. Add one gallon of enzyme and/or bleach if indicated on TPSS control panel.
3. If a used canister has sat for a long period of time and fluids have dried onto the sides, or if a canister has contents
which are very rich in whole blood, run a second cycle or long cycle. A long cycle is started by pushing the “WASH”
button twice within five seconds.
Problem: Only One Side Cleans
Possible Solution
1. Call DMS at 1-888-466-6633.
2. Continue to use operative side until repairs are made.
Problem: Unit Will Not Operate
Possible Solutions:
1. Check that the machine is plugged into an operational outlet.
2. Check and reset the ground fault switch for on the outlet.
3. Call maintenance to replace the Unit internal fuses.
4. Check that the TPSS lid is down and that canisters are locked in and registered on display.
5. Check for error display and review specific code for action. CALL DMS with six-digit error code.
Problem: The Unit’s Canister Rings Did Not Go Back to Home
Possible Solution:
Call DMS at 1-888-466-6633. Use the opposite side until service is complete.
Problem: Error Code Displayed
Possible Solutions:
1. Write down error code(s). For more than one error code, press the abort button until all error codes are displayed.
2. Call DMS at 1-888-466-6633 with error codes.
Problem: Water on Floor Around Unit
Possible Solution:
Check plumbing and back flow preventer for leaks. Call maintenance for repair.
Problem: Strong Bleach Odors
Possible Solutions:
1. Improve area ventilation.
2. When changing bleach bottle, replace lid of old bottle before disposing of bottle.
3. Call DMS at 1-888-466-6633.