![Dornoch TP-SS100 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/dornoch/tp-ss100/tp-ss100_technical-and-service-manual_2523230026.webp)
Safety Station Technical Service Manual
23
10/17/2011
GD-99009 Rev. B
D e -in s ta ll/Dis c on ne c t In s t ru c t ions
Please contact DMS at 1-888-466-6633 prior to de-installation or re-installation of TPSS units.
If a Customer de-installs or re-installs a TPSS, the guidelines listed below must be followed.
•
Always wear personal protective equipment (PPE), including eye protection and latex gloves, to prevent
contact with potentially infectious material.
•
Always unplug the unit from the wall electrical receptacle before moving equipment.
•
Wipe down the unit with an approved contact disinfectant after each use.
E q u ipm e nt S t or ag e
Temperature range for proper storage is -20
o
C to 40
o
C. If unit is to be taken out of service and stored for a prolonged
period of time, evacuate water from the hot and cold water manifolds, as well as removing the bleach and enzyme
solution from their chemical tanks regardless of temperature within storage facility.
Operation
Storage and Transportation
Temperature
Relative Humidity
Atmospheric Pressure
Figure 13 – Environmental Conditions