32
Latch Status
This allows a latching flag to be used as a trigger. The flag can be set or cleared
by any combination of the four outputs. See Config Set Latch on how to set and
clear this flag.
Off
The output or function ignores this trigger.
Hi
This trigger is true when the flag has been set.
Lo
This Trigger is true when the flag is cleared.
Return to Zero
This trigger is based upon the return to zero flag. The return to zero flag is set
any time the gross weight on the scale returns to gross zero band. The gross zero
band value is set by the Print Threshold parameter. When the return to zero
trigger is set to Yes or No, the status of the return to zero flag is checked and
used as a trigger. The return to zero flag is false when a standard auto print
occurs or when a trigger base print function occurs.
Off
The output or function ignores this trigger.
No
This trigger is true when the scale has not returned to zero after auto
printing.
Yes
The trigger is true when the scale has returned to gross zero and has not
met the print conditions.
Zero Band
This trigger is based upon the Print Threshold value. The trigger state is based
upon either being within or outside the Print Threshold.
Off
The output or function ignores this trigger.
In
This trigger is true when the weight display is within the Set Print
Thresh percentage of capacity.
Out
This trigger is true when the weight display is outside the Set Print
Thresh percentage of capacity.
Set Latch Config
This parameter allows one or more of the four outputs to set or clear the Latch
Flag. When enabled, the output will set or clear the Latch Flag as the output
switches from active to inactive. The Latch Flag will not change unless an
output is configured to do so by setting this parameter to Set or Clear.
Off
Does not alter the Latch Flag parameter above.
Set
When the output status transitions from active to inactive, the Latch
Flag is set.
Clear
When the output status transitions from inactive to active, the Latch
Flag is cleared.
Содержание GuardianXL
Страница 2: ...1...
Страница 8: ...7...
Страница 16: ...15...
Страница 18: ...17 Figure 3 1 Motherboard Load Cell Connection and Jumper Location...
Страница 22: ...21...
Страница 38: ...37 REMOVE THIS PAGE AND REPLACE WITH FIG4 7 PDF PRINT B SIZE Figure 4 7 Setup Mode Menu...
Страница 39: ...38...
Страница 55: ...54...
Страница 84: ...83...
Страница 86: ...85 Figure A 1 Indicator Dimensions Dimension A Standard GuardianXL 6 0 GuardianXL with Relay Option 6 5...