FMX Generator Directions for Use
7
• Before cleaning, unplug any power module that may be attached to
the generator, and unplug the generator from the power source.
• Do not immerse the generator or foot pedal in any liquid.
• Do not use an abrasive cloth or abrasive cleaner anywhere on the generator.
The generator, power cord and foot pedal may be cleaned
with a soft cloth dampened with water, alcohol, mild soaps,
detergents or disinfectants. Non-flammable agents should be
used for cleaning and disinfection wherever possible.
Generator Power-Down Instructions
To power-down the generator, press the front panel power button
and hold for 1 second. A short beep will sound, and the message
“Powering Down” will display on the LCD touchscreen.
When power-down is complete, the screen will go dark. It is now safe
to turn off the rear power button and disconnect the power cord.
Symbols
The following symbols are found on the front and back panels of the generator
as well as on the accompanying documents of the generator and accessories.
Consult Instructions For Use
Single-use Only
DO NOT USE if package appears damaged
Serial Number
Lot Number
Date of Manufacture
Type CF Applied Part
Warning: Dangerous Voltage
Non-ionizing Radiation
Temperature Limitation
Caution
Alternating Current
Conductor may be used to provide
potential equalization between the
equipment and the installation busbar
Specifications
Operating Environment Conditions
• Temp: 10° to 30° C(50° to 86° F)
• Humidity: 5% to 85% Non-condensing
Storage Conditions
• Temp: -20° to 65° C (-4° to 149° F)
• Humidity: 15% to 85% Non-condensing
Shipping Conditions
• Temp: -20° to 65° C (-4° to 149° F)
• Humidity: 15% to 85% Non-condensing
Moisture Protection
• Generator: IPX1
• Foot Pedal: IPX7 minimum
Generator Dimensions
• 14” x 4” x 12”
(35.6 cm x 10.2 cm x 30.5 cm)
Generator Weight
• Approximately 20 lbs. (9 kg)
Power Requirements
• Voltage: 100/110/220/230 Vac
• Current: 4/2A rms max
• Frequency: 50/60 Hz
Power Output
• 60W Max
Fuses
• For 100/110V: T 5A L 250V
• For 220V/230V: T 2.5A L 250V
Fundamental Output Frequency
• 40.68 MHz
Mode of Operation
• Intermittent Operation
10 sec. ON, 30 sec. OFF
Protection Against Electric Shock
• Class I
Applied Part Classification
• Type CF
Содержание FMX