Tab. 3: Reference and Declaration of Manufacturer - Electromagnetic Immunity
The PulsePen is suitable for use in the specified electromagnetic environment. The user of the PulsePen should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic Environment
Conducted RF
EN 61000-4-6
from 150 kHz to 80
MHz
Home Healthcare
Environment
3 V RMS outside the
ISM band, 6 V RMS in
the ISM and amateur
radio bands
Professional
Healthcare
Environment
3 V RMS outside the
ISM band, 6 V RMS in
the ISM band
Not applicable
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2,7 GHz
80% AM at 1kHz
3 V/m
80 MHz to 2,7 GHz
80% AM at 1kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the PulsePen, including cables, than the
minimum recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d= 1,2√P 150kHz to 80 MHz
d= 1,2√P 80MHz to 800 MHz
d= 2,3√P 800MHz to 2,5GHz
Where
P
is the maximum output rating of the
transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters
(m).
Field strengths for fixed RF transmitter, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Levels characteristic of a typical commercial /
hospital environment.
Interference may occur in the vicinity of
equipment marked with the following symbol:
18
User Manual
PulsePen
- V. 5.0_b