Tab. 2: Reference and Declaration of Manufacturer - Electromagnetic Immunity
The PulsePen is suitable for use in the specified electromagnetic environment. The user of the PulsePen should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
IEC 60601-1-2
Test level
Compliance level
Electromagnetic Environment
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+/- 8KV contact
+/- 2KV, +/- 4KV,
+/- 8KV, +/- 15KV air
+/- 8KV contact
+/- 2KV, +/- 4KV,
+/- 8KV, +/- 15KV
air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
±2 KV
Frequency 100 kHz
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
+/- 0.5KV, +/- 1KV
differential mode
+/- 0.5KV, +/- 1KV, +/-
2KV common mode
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
Interruptions
and Voltage
variations on power
supply input lines
IEC 61000-4-11
Voltage dips
0% U
T
, 0.5 cycles
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% U
T
, 1 cycle /
70% U
T
, 25/30 cycles,
Single phase @ 0°
Voltage interruptions
0% U
T
, 250/300 cycles
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the PulsePen requires continued
operation during power mains interruptions, it
is recommended that the computer used with
the PulsePen be powered from an
uninterruptible power supply or a battery.
Magnetic Field at
Power frequency
(50/60 Hz)
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
17
User Manual
PulsePen
- V. 5.0_b