45
Separator
’;’ (ASCII=59)
1
Result field: g/l
pppp,pp or ’ --- ’ if not
defined
7
Separator
’|’ (ASCII=124)
1
Error code
ddd
(Note 4.)
3
Carriage return
(ASCII=13)
1
New Line
(ASCII=10)
1
End message
ETX (ASCII=03)
1
Note 1.:
Where ’TTTTT’ one of the following:
„APC „,”PROTC”,”PROTS”,”LA „,”PT „,”APTT „,”FIB „,”TT „,”D-DIM”,”A-
TIII”
„Neph „,”Turb „, „II „,”V „,”VII „,”X „ „VIII „,”IX „,”XI „,”XII „, „QC
„,”undef”
Note 2.:
Where C: measuring position ( ’1- 4’), ssss,s: secundum (’ 123,4’)
In the parallel measuring case the position number of „Raw result1” and
„Raw result2” will be different.
In the simple mode position number and secundum will be equal. (Raw
result 1 = Raw result
Note 3.:
The result is ’µgFEU/ml’ in case of DDIMER, „%” in case of ATIII.
Note 4.:
The „Error code” is the 3 digit decimal format of the error byte.
The bits are the following:
00 0000 0000: No error
00 0000 0001: Calibration error
00 0000 0010: Difference error during parallel measuring
00 0000 0100: External light too much
00 0000 1000: Curve error (suspicious data during measure phase)
00 0001 0000: Out of range error (time is out of normal range)
00 0010 0000: Incubation error (the time was too long)
00 0100 0000: Expired lot
00 1000 0000: Control result out of limit
01 0000 0000: Reagent control differences
10 0000 0000: There is no derivate fibrinogen calibration data