20
7. Installation of the Coag 2D
7.1 Installation Requirements
Place the instrument on a horizontal laboratory table. It requires an empty
space at least 300x300mm in size.
Place the instrument near a grounded mains power connector that has a
maximum of 1.5 A input current.
Place the instrument in a way that the power cord can be easily
disconnected if necessary. Do not place the instrument in direct sunlight.
7.2 Connecting to the Electrical Network
Connect the DC cable of the power unit to the socket on the rear of the
device.
7.3 Testing the Functionality of the Instrument
Turn on the instrument and check whether the main menu appears after
the software has finished loading. Check the temperature displayed, then
wait until it turns green. Note that the instrument will only allow measure-
ments within the prescribed temperature range (37°C + 0.1).
Figure 7.1 - Main menu