semi-automated coagulometer
User Manual
Reference No.: 11000102EN
Internal Reference No.: DIAB102AEN
Страница 1: ...semi automated coagulometer User Manual Reference No 11000102EN Internal Reference No DIAB102AEN...
Страница 2: ...ument applies to the latest software version listed and higher versions When a subsequent software version changes the information in this document a new electronic edition is released and supplied by...
Страница 3: ...2 Specifications 17 6 Components of the Coag 2D 18 6 1 Structure of the Instrument 18 7 Installation of the Coag 2D 20 7 1 Installation Requirements 20 7 2 Connecting to the Electrical Network 20 7 3...
Страница 4: ...es 37 13 2 Control Measurement 38 13 3 Processing Control Results 38 14 Maintenance of the Coag 2D 39 14 1 Cleaning the Measuring Unit 39 14 2 Cleaning the Screen 39 14 3 Cleaning the Housing 39 14 4...
Страница 5: ...lotting time performing chromogenic and immunological assays on plasma specimens prepared from sodium citrate anticoagulated blood samples collected into standard sample tubes For the best Coag 2D exp...
Страница 6: ...r spoken between the user and Diagon Ltd does not constitute as a confidential relationship Such information is considered non confidential which means Diagon Ltd may reproduce publish or use it witho...
Страница 7: ...2 Package and Accessories Make sure all items listed are present in the package when unboxing Coag 2D analyzer 1 pc Power supply unit 1 pc Magnetic mixer 2 pcs set 1 set User Manual this document 1 p...
Страница 8: ...able Symbols 3 2 Symbols on Accessories Serial Number Catalogue Number In Vitro Diagnostic Device CE Conformity Read the Manual Manufacturer information Requires specific disposal Single use only Non...
Страница 9: ...9 3 3 Symbols on the Packaging Fragile Keep Dry Proper Storage Suitable Temperature Range Suitable Humidity Range...
Страница 10: ...r the list of manufacturer approved consumables visit the official Diagon website www diagon com Compliance with this manual is also part of the conditions of intended use make sure to follow these in...
Страница 11: ...in a room with good ventilation Make sure that there is no device that emits electromagnetic radiation or electronic interference in the proximity of the Coag 2D and avoid placing it near electronic w...
Страница 12: ...ling blood samples as they are considered biohazardous The controls and the calibrator are considered potentially infectious Always take the necessary protective measures when using them The liquid an...
Страница 13: ...sure the reagents do not come into contact with skin or soft tissue If the reagent comes into contact with your eyes rinse them with plenty of water and then seek medical aid Do not ingest reagents if...
Страница 14: ...rinciple 5 1 1 Clotting Time Determination The blood plasma and the reagent are mixed in a plastic cuvette with the appropriate timing The cuvette is placed in the measurement position located in the...
Страница 15: ...oftware of the instrument places a calibration curve on the input function points and based on the curve unknown concentration or percentage values can be determined based on the measured coagulation...
Страница 16: ...function points and based on the curve unknown concentration values can be determined from the measured light intensity change The calibration curve points may be provided by the manufacturer of the r...
Страница 17: ...aces a calibration curve on the input function points and based on the curve unknown relative concentration values can be determined from the measured light intensity change The points of the calibrat...
Страница 18: ...block regulated to 37 C It has four pre heated cuvette positions two measure ment locations and two reagent positions for reagent vessels Reagent Positions There is a rotating magnetic field at the r...
Страница 19: ...like other peripherals USB storage devices may be connected while the instrument is on however a 5 10 second waiting period is required before any action may be performed on them Slave USB connector f...
Страница 20: ...y disconnected if necessary Do not place the instrument in direct sunlight 7 2 Connecting to the Electrical Network Connect the DC cable of the power unit to the socket on the rear of the device 7 3 T...
Страница 21: ...nctional by tapping any button on screen The content displayed depends on the item selected To return to the main menu tap the Esc button Check whether the date and time displayed are correct In case...
Страница 22: ...arameter turns orange To select a diffe rent parameter tap a different line To adjust the parameter values tap the pencil icon A virtual keyboard appears its button layout depends on the parameter sel...
Страница 23: ...t stops the measure ment and the result will be marked with a T flag Lag Time During this time the software disregards any changes in the cuvette Reagent Nr Identifies the number of reagents Incubatio...
Страница 24: ...etween the preparation of the reaction mixture and the endpoint of the reaction Reagent Nr Identifies the number of reagents Incubation Time Sets the time period between adding the sample and adding t...
Страница 25: ...lable Options Language English Polish Ukrainian French Spanish Italian Date Time Date Time Settings Sound On Off Printer On Off Export Import settings In order to prevent data loss data can be exporte...
Страница 26: ...rument see section 6 1 Note it is recommended to perform this action before upgrading the instrument software as all settings will be erased by resetting the instru ment 9 2 Error Messages The error f...
Страница 27: ...when dispensing samples the tip has to be replaced between each sample to prevent carryover Only use reagents controls calibrators and consumables that are manufactured accordingly with measurement r...
Страница 28: ...surement win dows are displayed FIgure 10 1 Measurement windows To start the measurement tap the upper part above the line of either measurement windows Figure 10 2 Entering sample ID A virtual keyboa...
Страница 29: ...he required action from the user The system always displays the next required action in the status row After entering sample ID the first action is Add Cuvette Depending on the test setup of the actua...
Страница 30: ...nnel s Zero green Add Sample 4 Sample added according to the pro tocol described in the appendix Incubation time oran ge Incubating 5 Incubation time is over sound signal Incubation time oran ge Add R...
Страница 31: ...rcode contains three lines consisting of 16 alphanumeric characters The software can only correctly interpret the barcode system used by Diagon In case control material is entered and there is a range...
Страница 32: ...code in this menu item the barcode can be entered either by typing it in manually or by using the barcode reader which is connected to the USB port The data can be set and validated by tapping the tic...
Страница 33: ...percentage values the right column contains the OD min values These values of the calib ration curve can be recorded by measuring specific calibrators 11 4 Producing a Dilution Series The reagent man...
Страница 34: ...ata of 1000 patients in the format shown in Figure 12 1 Tap the Result button to access this menu Figure 12 1 Results Menu The test results are displayed in chronological order To move back and forth...
Страница 35: ...en The test selected will be highlighted with an orange background The original results calculated values measurement channel number the possible error codes and whether or not a particular test has b...
Страница 36: ...rial have to be mea sured and registered by the user It is recommended to use QC material delivered by Diagon or its distributors The instrument software stores the results of QC measurements the meas...
Страница 37: ...ch is located on the top of the screen Target values may be set for four different tests per control material QC data may be entered manually based on the QC data sheet provided or by reading them in...
Страница 38: ...identical to a regular mea surement procedure except when entering the ID previously defined QC materials may be chosen The results are stored in the QC result databa se which is accessible from QC m...
Страница 39: ...h cotton buds soaked in water mixed with a washing agent Leave the unit to dry in heated position for at least ten minutes after clea ning and before starting measurements 14 2 Cleaning the Screen Cle...
Страница 40: ...f measurement Check with standard PT assay and nor mal sample Occurs with all assays Frozen software Switch off and restart Repeated freezing Temperature regulation failure Switch off Call service 14...
Страница 41: ...oes not start in time because the incubation time is too long O Incubation was overheated If the thermostatic block is too warm E Expired lot The reagent used has expired R Barcode type error If the r...
Страница 42: ...37 C Dia FIB at room temperature Dia TT at room temperature Sample pre paration Plasma control 1 10 dilution with Dia Imi dazol Sample dosing 50 l plas ma control 50 l plas ma control 50 l Dia PTT 10...
Страница 43: ...Sample preparation Plasma control 1 20 dilution with Dia ATIII FIIa Diluent Sample dosing 20 l plasma control 115 l Dia Ddi Buffer 50 l prepared sample 50 l Dia ATIII FIIa Incubation 2 minutes 2 minu...
Страница 44: ...or ASCII 124 1 Time HH MM 5 Separator ASCII 124 1 Measuring type TTTTT Note 1 5 Separator ASCII 124 1 Raw result 1 C ssss s Note 2 8 Separator ASCII 124 1 Raw result 2 C ssss s Note 2 8 Separator ASCI...
Страница 45: ...number and secundum will be equal Raw result 1 Raw result Note 3 The result is gFEU ml in case of DDIMER in case of ATIII Note 4 The Error code is the 3 digit decimal format of the error byte The bit...
Страница 46: ...46 e g Expired lot and the Curve error in same time 01001000 hex 48h the error code is 072...
Страница 47: ...removing and reformatting the pendrive to a FAT32 file system Copy the update files back onto pendrive and attempt the update again 7 When firmware update is finished you need to press the Reboot butt...
Страница 48: ...48 13 Answer Yes when the instrument prompts so 14 Switch the coagulometer off when the process is finished 15 The instrument software is now up to date and ready to use from the next restart...