- Troubleshooting
12.2 Error codes and solutions
When the instrument detects an operating error, a letter will appear on the top left corner of the display. In
M00
,
M01, M02
,
M03
,
M90
and
M08
can be displayed the error message. Refer to the following table for the solution:
Error
codes
Displayed message
Cause
Solution
R
System Normal
No error
I
Detected No Signal
1) No detected signal 2) Sensors installed improperly 3)
Too many encrustations
4) Pipe lining too thick 5)
The transducer cables are not properly
connected
1) T
ight the transducers on the pipe
2) Check the setted parameters 3)
Clean the pipe surface and remove any rust
4) Change installing position
5) W
ait until the new liner is dry
6) Check the cables.
J
Hardware Error
Hardware problems
Contact headquarters
H
PoorSig Detected
1) W
eak signal detected
2) Sensors installed improperly 3)
Too many encrustations
4) Pipe lining too thick 5)
The transducer cables are not properly
connected
1) T
ight the transducers on the pipe
2) Check the setted parameters 3)
Clean the pipe surface and remove any rust
4) Change installing position
5) W
ait until the new liner is dry
6) Check the cables.
E
Current Loop over 20mA
1) 4÷20mA
current loop over 100%
2) Current output Improper settings
Check the values setted in menu M56 and if
the actual flow is too high
Q
Frequ OutputOver
1) Frequency output over 120% 2) frequency output Improper setting or actual flow too high
Check the values setted in menu M66-M67-
M68 and M69 and if the actual flow is too
high
F
Refer to table 2
1) Error in self-diagnosis during power-on 2) Permanent hardware error
1)
Restart
2) Contact headquarters
G
Adjustig gain S1-S2-S3-S4 (displayed on M00, M01,M02 and M03)
The
instrument
is
running
the
automatic
checkup
and
the
numbers
indicate
the
sequential
progress
K
Empty pipe
1) No liquid inside the pipe 2) Setting error in M29
1) Reposition the transducer where the pipe is
filled with fluid
2) Set 0 in M29
Tab. 13