DI MPU01 Скачать руководство пользователя страница 17

Pag. 17 of 52

- Installation

4)  Proceed to the electrical connection

 7.7 Installation check-up

The signal reception power and quality (

Q

) can be checked through the installation check-up and it’s possible to make 

the comparison of the measured flight time by the measuring range depending on the pipe diameter. 

7.7.1 -  

Signal reception power - M90

The signal reception power, displayed to the menu 

M90 

with

 UP 

and

 DN

, is indicated by a three-digit number. 

[00.0] means missing signal and [99.9] indicates the maximum measurable value.  

Although the instrument is working properly with a signal power between 50.0 and 99.9, it is always recommended 

to try to get a value as high as possible using the following methods:

  (1) Select the most favorable installation position.

  (2) Clean the pipe outer surface and apply more coupler grease. 

  (3) Move the transducers both vertically and horizontally while the signal reception control.

   

Mechanically lock the transducers when the detected power has reached its maximum value (always checking  

   

that the distance between the two transducers is equal to that indicated in the menu 

M25

)

7.7.2 - Signal quality (Q) - M90

Better the signal quality (

Q

 value higher), better will be the ‘SNR and consequently the accuracy. In normal operating 

conditions the 

Q

 value, displayed to the menu 

M90

, is between 60 and 90. In the case of a lower value, check:

  (1)  Any interference with other instruments.

  (2) The transducers coupling with the pipe surface ube (clean the pipe or add more coupler grease) 

  (3)  The mounting position on the pipe

7.7.3 - Total transit time and Delta Time - M93

The numbers shown in the 

M93

 are called total transit time and delta time. These values   are fundamental to calculate 

the flow rate inside the pipe. 

The total transit time should remain stable or in any case subject to minimal variations. 

If the delta time fluctuates above 20% means that there are problems with the transducers installation. 

Fig. 32

3) the inclined surfaces of the ultrasonic transducer 

  signal emitter (1 in figura 27) should be directed

  towards each other

Fig. 31-a

Содержание MPU01

Страница 1: ...r 0 20mA Accuracy 1 Repeatability 0 2 Linearity 0 5 Basic measurement period 500ms Serial port RS485 Communication protocol MODBUS RTU or ASCII opt Programmable frequency output 0 5000Hz Relay output...

Страница 2: ...n or at Z mode 1 sonic section The installation method choice depends on the pipe and the fluid characteristics The instrument measures the transit time via the two transducers that alternatively tran...

Страница 3: ...e flow rate measurement and separate totalization Data Converter Unit Accuracy 1 Working conditions Temperature 20 85 C humidity 85 non condensing for applications in conditions different from the sta...

Страница 4: ...perature 30 90 C Menu 23 19 Clamp On TS 2 TS 2 HT Dimensions Pipe range 20 100mm 4 Temperature 30 160 C Menu 23 19 Clamp On TS 2 TM 1 Dimensions Pipe range 50 700mm 2 40 Temperature 30 90 C Menu 23 16...

Страница 5: ...re 30 900 C Menu 23 20 Clamp On TL 1 TC 1 standard Dimensions Pipe range 80 4000mm 3 160 Temperature 40 400 C Max pressure 1 6Mpa 16bar Menu 23 17 Insertion TC 1 TLC 2 for non metallic pipes Dimension...

Страница 6: ...ower cables from the DIMPU connection cables 2 Remove the caps from the cable glands and open the cover by unscrewing the screws 3 Lead the cables into the transmitter through the cable glands 4 Close...

Страница 7: ...rical connections use a cable with a 6 10mm outer diameter and fully tighten the M18 cable gland Securely close the cover position the cable so that it forms a downward curve at the M18 output Fig 4 i...

Страница 8: ...Pag 8 of 52 5 4 TRANSDUCER CONNECTION Electrical connections 5 3 2 Supply voltage in 10 30Vdc Fig 6 Fig 7...

Страница 9: ...Pag 9 of 52 Electrical connections 5 5 OUTPUT SIGNALS CONNECTION 5 5 1 Analog output 5 5 2 Pulse output Fig 8 Fig 9...

Страница 10: ...Pag 10 of 52 Electrical connections 5 5 3 Relay output 5 5 3 MODBUS port Fig 10 Fig 11...

Страница 11: ...ion All configurations set by the user are stored in memory but it s good to make sure that the menu M26 Default Settings is set to 0 Use RAM Settings 6 2 keyboard DIMPU has 4 buttons Press to activat...

Страница 12: ...the flow velocity in the pipe One of the best devices is the transducers upstream installation of a fluid threads rectifier which allows to have a straight section length of the pipe less than indica...

Страница 13: ...hey may not be completely filled with fluid fig 16 inclined pipes with the downward flow because they may not be completely filled with fluid fig 16 the transducers must never be placed in the highest...

Страница 14: ...pipes with diameters in the DN50 250 range Installation 7 4 Z installing Is the installation method for pipes with diameters in the DN300 400 range 7 5 W installing Is the installation method for pip...

Страница 15: ...the menu M23 choose the option 17 or 21 TC 1 or TLC 2 insertion transducer in the menu M24 choose 1 Z installation and in the menu M25 check the positioning distance 5 Determine the best location for...

Страница 16: ...lose the ball valve to avoid leakage 6 Use a meter or a caliper to measure the A dimension Slowly insert the transducer into the valve support and open the ball valve Measure the distance L between th...

Страница 17: ...ck the transducers when the detected power has reached its maximum value always checking that the distance between the two transducers is equal to that indicated in the menu M25 7 7 2 Signal quality Q...

Страница 18: ...uge 8 3 Pipe material Value detectable on site or from the technical data of the hydraulic interested programming in M14 8 4 Materiale del rivestimento interno del tubo Value detectable on site or fro...

Страница 19: ...9 3 Marking tools To trace on the pipe surface the transducers positioning points are sufficient simple tools but effective at the same time a paper roll piece like that calculators with a width grea...

Страница 20: ...oint to a length equal to half the r circumference Next at 180 degrees on the circumference r from point a draw a straight line called D parallel to the straight line S and with equal length The inter...

Страница 21: ...the transducer on the pipe surface at the transducer installation point already cleaned 3 Securely fasten with a metal fixing clamp or other the transducer on the pipe WARNING do not overtighten to a...

Страница 22: ...ogramming M76 2 Q max alarm programming M24 Transducers mounting method programming M77 Buzzer operation programming M25 Transducers mounting distance displaying M78 OCT output programming M26 Data st...

Страница 23: ...w totalizer measurement unit 10 2 4 How to enable and disable the totalizers Use the M34 M35 and M36 menu to enable and disable the forward POS reverse NEG or net NET flow totalizer 10 2 5 How to rese...

Страница 24: ...ized volume can be sent as an output pulse The totalizer will generate one pulse per volume unit The pulse totalizer can be generated by the OCT relays or BUZZER hardware devices For example configure...

Страница 25: ...change will be saved in EEPROM 10 2 19 How to set the RS485 serial interface Use M62 for setting 10 2 20 How to display the partial totalizers Use M82 to display the partial totalizer daily monthly o...

Страница 26: ...rd totalizer POS value 11 03 M03 Instantaneous flow rate and reverse tot displaying Displaying only The display shows the instantaneous flow rate and reverse totalizer NEG value 11 04 M04 Instantaneou...

Страница 27: ...s the previously set value By entering a new value the system will automatically calculate the pipe inner diameter new value M13 11 13 M13 Pipe inner diameter programming The display shows the previou...

Страница 28: ...shows the previously set value Available only with 11 option selected in M16 11 18 M18 Pipe inner lining thickness programming The display shows the previously set value 11 19 M19 Inner ABS thickness...

Страница 29: ...characteristics when 3 option is selected 4 Standard B 5 Insertion B 45 6 Standard L 7 JH Polysonics 8 Standard HS 9 Standard HM 10 Standard M1 11 Standard S1 12 Standard L1 13 PI Type 14 FS410 FUJI 1...

Страница 30: ...and other configuration parameters aren t saved or modified 11 27 1 Default setting loading To load a default setting proceed as follows press Select the default setting to be loaded with or and pres...

Страница 31: ...neous flow rate measurement unit proceed as follows press Press or to select the volume measure unit and press to confirm Available settings Cubic Meter m3 Liter l US Gallon Gal UK Gallon IGL Million...

Страница 32: ...rmation in 2 distinct sub menu is needed It is also possible to reset all totalizer or single totalizer To reset proceed as follows press Select with or YES and press Selecting NO the reset procedure...

Страница 33: ...ibration By pressing during the calibration procedure is stopped The digit at the bottom left indicates the remaining reads number for the calibration completion NB During calibration the signal statu...

Страница 34: ...to confirm and unlock the programming changes Programming menu 11 48 M48 Calibration data programming Only for headquarters use 11 49 M49 MODBUS serial port test Only for headquarters use 11 50 M50 Da...

Страница 35: ...move the cursor to the right with Press to store the data logger interval time Set the data logger recordings number to be made Change the digit with and move the cursor to the right with Press to sto...

Страница 36: ...et to 0 4 20mA the measure unit is m3 h with M55 set to 3 4 20mA vs Vel the measure unit is m s 11 58 M58 Analog output simulation In this menu it s possible to force the analog output signal value to...

Страница 37: ...ramming In this menu it s possible to set theAL4 analog input beginning and full scale 11 66 M66 AL5 analog input programming In this menu it s possible to set theAL5 analog input beginning and full s...

Страница 38: ...et the maximum flow threshold for the 1 alarm 11 75 M75 2 Q min alarm programming In this menu it s possible to set the minimum flow threshold for the 2 alarm 11 76 M76 2 Q max alarm programming In th...

Страница 39: ...ramming In this menu it s possible to set the function associated to the relay output Available settings 0 No Signal 1 Poor Signal 2 Not Ready No R 3 Reverse Flow 4 AO Over 100 5 Fo Over 120 6 Alarm 1...

Страница 40: ...ossible to display the DIMPU recorded daily monthly and annual events To display the events proceed as follows press Select the events display mode with or Available settings 0 Browse by Day 1 Browse...

Страница 41: ...alue x1 Available settings 0 x0 0001 E 4 1 x0 001 1E 3 2 x0 01 3 x0 1 4 x1 5 x10 6 x100 7 x1000 8 x10000 E4 9 x100000 E5 10 x1000000 E6 11 89 M89 Temperature differential displaying In this menu it s...

Страница 42: ...alues 11 93 M93 Flight time and delta T In this menu it s possible to display the flight time measured by the DIMPU and the difference in flight times UP DN 11 94 M94 Reynolds number displaying In thi...

Страница 43: ...nd press to confirm in the example shown next the PT100 Temperature function Available operations x 1 x abs x x x sqrt x exp x ln x log x power x y sin x cos x arcsin x arccos x arctan x Store in M x...

Страница 44: ...Pipe Material Carbon Steel Stainless Steel Cast Iron Ductile Iron Copper PVC Plastics in general Aluminium Asbestos Fiberglass Epoxy Other M14 Inner lining material Liner Material None No Liner Tar Ep...

Страница 45: ...material eventual inner lining and its thickness transducers mounting method M25 Instantaneous flow rate measure unit Flow Rate Unit Measure units associated with the instantaneous flow rate measurem...

Страница 46: ...the measured flow velocity is less than the cutoff value the display will show the instantaneous flow rate measure at fixed 0 Range 0 000 0 25m s M41 Zero flow calibration Set Zero When the fluid in...

Страница 47: ...utput Value Is the instantaneous flow rate value expressed in the above selected measure unit which is associated with the analog output scale beginning 4 or 0mA M56 4 20mA output full scale VALORE RI...

Страница 48: ...rameter data error Press ENT key and restore default parameters System Data Memory Error System stored data block error Restart or contact headquarters Circuit Hardware Error Sub CPU circuit fatal err...

Страница 49: ...tations 4 Pipe lining too thick 5 The transducer cables are not properly connected 1 Tight the transducers on the pipe 2 Check the setted parameters 3 Clean the pipe surface and remove any rust 4 Chan...

Страница 50: ...hour d ddddddE dd CR LF DQM CR Return flow rate per minute d ddddddE dd CR LF DQS CR Return flow rate per second d ddddddE dd CR LF DV CR Return flow velocity d ddddddE dd CR LF DI CR Return POS tota...

Страница 51: ...ll be W12345DV CR 13 3 3 N prefix The prefix N is a single byte IDN network prefix not recommended in a new design It is reserved only for the purpose of the compatibility with the former versions Com...

Страница 52: ...Documentation subject to technical change with no prior warning...

Отзывы: