11.26 M26 - Data storage mode programming
The display shows the previous setting.
Available settings:
0. Use RAM Settings
(recommended); all the latest programming
changes are automatically stored in the EEPROM
1. SALVA PAR. IMPOSTAZIONI; to store in the EEPROM the latest
changes to any menu, must go
back to
M26
and press
- Programming menu
11.27 M27 - Default settings library
The display shows the previous setting.
Here it is possible save or load the flow measurement default settings, from
M10
to
M24
, previously stored on EEPROM, (es. pipe diameter, thickness, etc.).
Outputs and other configuration parameters aren’t saved or modified.
11.27.1 - Default setting loading
To load a default setting, proceed as follows:
press
Select the default setting to be loaded with
or
and press
Confirm the selection by pressing
. The system automatically displays
M23
11.27.1 - Storing default setting
To store a new default setting, proceed as follows:
press
Select with
or
to replace the default setting and press
To select the “
1.
” option, press
or
and confirm by pressing
.
The system automatically displays
M23
11.28 M28 - HOLD mode programming
The display shows the previous setting.
Selecting “
YES
”, in the case of transducer echo signal temporary loss, the
DIMPU
maintains the last valid measurement
Selecting “
NO
”, the instantaneous flow rate measured value will go to zero