PPCK+™ Operation and Maintenance Manual
Page 87
©2001
DH Instruments, Inc
.
MODE
Purpose
Read or set the control mode.
Command
Query
“MODE
n
”
“MODE?”
Classic “MODE=
n
”
“MODE”
Arguments
n:
’0’ for static pressure control.
’1’ for dynamic pressure control.
Remarks
The method which the PPCK+ controls pressure is selected with the MODE
program message. When the control mode is set, control parameters go to default
parameters for that range. The mode setting is range dependent.
Example
(enhanced)
Cmd sent:
Query reply:
“MODE 1”
“1”
Example
(classic)
Sent:
Reply:
“MODE=1”
“MODE=1”
Errors
ERR# 6:
The argument is invalid.
See Also
3.1.2.4, 3.2.3, 3.2.6
MSGFMT
Purpose
Read or set the type of program message format to use.
Command
Query
“MSGFMT
n
“MSGFMT?”
Classic “MSGFMT=
n
“MSGFMT”
Arguments
n:
‘1’ to use the enhanced command format.
‘0’ to use the classic command format.
Defaults “MSGFMT
0”
Remarks
The user can select the type of remote command format to use. This format must
agree with the format sent to the PPCK+.
Example
(enhanced)
Cmd sent:
Query reply:
“MSGFMT 1”
“1”
Example
(classic)
Sent:
Reply:
“MSGFMT=1”
“MSGFMT=1”
Errors
ERR# 6:
Missing or improper program message argument(s).
See Also
4.3
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...