PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 50
OPERATION
The AutoZ function and values are range AND mode (gauge or absolute) specific.
To access the PPCK+ AutoZ function press
[SPECIAL]
and select
<1AutoZ>
. The display is:
Selecting
<1off> (or <1on>)
is for changing the AutoZ status for the current range and
measurement mode from ON to OFF or vice versa. Selecting
<1off>
turns OFF the
AutoZ function for the current range and mode (see AutoZ ON/OFF of this Section).
AutoZ ON is indicated by a <z> in the main run screen, top line, third character from
the right. When AutoZ is OFF, the character is blank.
Selecting
<2view>
allows the current value of ZOFFSET for the current range and mode
to be viewed.
Selecting
<3edit>
allows the value of ZOFFSET for the current range and measurement
mode to be edited.
Selecting
<4run>
allows the routine for determination and activation of ZOFFSET by
measurement of ZSTD to be executed.
3.4.1.1
VIEW AUTOZ
To view the current ZOFFSET for the current range and measurement mode press
[SPECIAL]
and select
<1AutoZ>
,
<2view>
. The display is:
1. Current measurement mode.
2. Current
range.
3. Display field of current value of ZOFFSET.
ZOFFSET should be zero when the PPCK+ is new or has just been calibrated.
ZOFFSET in absolute mode should be roughly equal to atmospheric pressure for an
absolute RPT operating in gauge mode.
The value of ZOFFSET is always displayed and entered in Pascal (Pa) as PPCK+'s
internal operations are in Pa.
1off 2view H3
3edit 4run
2
3
ZOFFSET absolute H3
0.0 Pa
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...